FDA Medical Device Recalls (openFDA)

Mermaid Medical A/S: The D*Clot HD Rotational Thrombectomy System is a percutaneous catheter-based system available in 6 French. The disposable system consists of an outer sheath through which a rotatable radiopaque spiral shaft extends ending in an eccentric blunt distal tip. Once activated, the shaft rotates at approximately 10.000 RPMs to break thrombus into smaller particles, which can be simultaneously aspirated using a dedicated vacuum syringe. The catheter is a stand-alone system and is powered by a 4.5 VDC alkaline battery pack in the control unit.

Recall number
Z-2213-2024
Recalling firm
Mermaid Medical A/S
Product
The D*Clot HD Rotational Thrombectomy System is a percutaneous catheter-based system available in 6 French. The disposable system consists of an outer sheath through which a rotatable radiopaque spiral shaft extends ending in an eccentric blunt distal tip. Once activated, the shaft rotates at approximately 10.000 RPMs to break thrombus into smaller particles, which can be simultaneously aspirated using a dedicated vacuum syringe. The catheter is a stand-alone system and is powered by a 4.5 VDC alkaline battery pack in the control unit.
Status
Open, Classified
Initiated
2024-05-01
Posted
2024-06-26
Terminated
not on file
Reason
Internal product testing failed bubble leak testing (ASTM F2096), which indicates the sterility barrier may not be fully intact.
Root cause
Under Investigation by firm
Action
The consignees and distributors were sent out the notification via email on 05/01/2024. The letter instructs the consignees to: 1. Quarantine and Discontinue Use of Affected Product: Identify, segregate, and discontinue use of the affected product in your inventory. 2. Complete and Return the attached Response Form: Complete the attached Response Form and return to Mermaid Medical by May 10, 2024 via email or by post. 3. Distributors/ If You Further Distributed Affected Product: Forward this communication to any customers that were distributed affected product. Instruct any of your customers to complete the response form and return to Mermaid Medical. Once they complete and return the response form, Mermaid Medical will provide instructions on how to return or destroy any affected product remaining in their inventory.
Quantity
190 devices
Distribution
Worldwide - US Nationwide distribution in the states of OH, MO, CO, NC, TX, MA, GA, FL, IL, IN, MD, PA, VA, and the country of Chile.
Lot, serial or model codes
Unique Device Identifier: 15711055004333; Lot codes: S2306001; Model Number: T10106065
510(k) numbers
["K021641"]
PMA numbers
not on file
Product code
QEW
Firm FEI
3007146453
Firm city
Stenlose
Firm state
not on file
Firm country
Denmark

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register