CareFusion 203, Inc.: Pulmonetic Systems P/N 19189-001 Patient Circuit SPU W/O PEEP 15 FT 5/Pkg. These circuits are intended for use with the MR Conditional LTV 1200 System with the ventilator utilizing an external exhalation valve, as single patient use (SPU) only patient circuits for adult and pediatric patients. The LTV 1200 ventilator is intended to provide continuous or intermittent ventilatory support for the care of individuals who require mechanical ventilation. The ventilator is suitable for use in institutional, home and transport settings, and is applicable for adult and pediatric patients weighing at least 5 kg (11 lbs.), who require the following types of ventilatory support: -Positive Pressure Ventilation, delivered invasively (via ET tube) or non-invasively (via mask). -Assist/Control, SIMV, CPAP or NPPV modes of ventilation. -Breath types including Volume, Pressure Control and Pressure Support.

Recall number
Z-2205-2010
Recalling firm
CareFusion 203, Inc.
Product
Pulmonetic Systems P/N 19189-001 Patient Circuit SPU W/O PEEP 15 FT 5/Pkg. These circuits are intended for use with the MR Conditional LTV 1200 System with the ventilator utilizing an external exhalation valve, as single patient use (SPU) only patient circuits for adult and pediatric patients. The LTV 1200 ventilator is intended to provide continuous or intermittent ventilatory support for the care of individuals who require mechanical ventilation. The ventilator is suitable for use in institutional, home and transport settings, and is applicable for adult and pediatric patients weighing at least 5 kg (11 lbs.), who require the following types of ventilatory support: -Positive Pressure Ventilation, delivered invasively (via ET tube) or non-invasively (via mask). -Assist/Control, SIMV, CPAP or NPPV modes of ventilation. -Breath types including Volume, Pressure Control and Pressure Support.
Status
Terminated
Initiated
2010-07-02
Posted
2010-08-10
Terminated
2011-12-17
Reason
CareFusion is recalling the P/N 19189-001 Patient Circuit (SPU W/O PEEP 15ft) because some of the circuit sense lines may have been reversed during the manufacturing process. If the sense lines have been reversed, immediately upon connection of the patient circuit to the ventilator and turning the ventilator on, the ventilator will begin to autocycle (with increased pressure delivered) with both
Root cause
Process control
Action
Customers were sent a CareFusion "Urgent Medical Device Recall Notification" dated July 2, 1010. The letter was addressed to Respiratory Care Manger. The letter described the product , problem and advised consignees to remove the affected product from inventory and promptly return it to CareFusion for replacement. Please contact CareFusion Technical Support at 800-754-1914, option 2, upon receipt of this notification to arrange for prompt return and replacement of any the patient circuits in your inventory.
Quantity
200 boxes (5 each)
Distribution
Nationwide distribution: AL, CA, FL, GA, HI, IL, IN, KS, MA,MI, MN, MT, NJ, NY, NC, OR, PA, RI, TN, TX, UT, VA, WA, and WI.
Lot, serial or model codes
Y09D1922 Y09H0564 Y09H0565 Y09K0491 Y09K0492 Y09S1283 Y09S1284 Y10A1230
510(k) numbers
["K083688"]
PMA numbers
not on file
Product code
CBK
Firm FEI
3002617992
Firm city
Minneapolis
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register