FDA Medical Device Recalls (openFDA)
Intuitive Surgical, Inc.: da Vinci S Surgical System, Model number IS2000 Endoscopic Instrument Control System, manufactured by Intuitive Surgical, Sunnyvale, CA.
- Recall number
- Z-2204-2008
- Recalling firm
- Intuitive Surgical, Inc.
- Product
- da Vinci S Surgical System, Model number IS2000 Endoscopic Instrument Control System, manufactured by Intuitive Surgical, Sunnyvale, CA.
- Status
- Terminated
- Initiated
- 2008-06-01
- Posted
- 2008-09-16
- Terminated
- 2009-07-29
- Reason
- Defective software chip may cause the system to fail and lock up.
- Root cause
- Software design (manufacturing process)
- Action
- Recall initiated June 1, 2008. Urgent Device Recall Letters were sent via e-mail, with follow up phone calls and service visits to correct the product. The firm recommended that continued use of the da Vinci S Surgical System during the time period prior to our service update to replace the software chip. The letter emphasizes the importance of always having backup equipment and instrumentation available when performing surgery with the da Vinci S Surgical System and being prepared to convert to alternative surgical techniques, if necessary. Customers are asked to fill out and return the acknowledgement form.
- Quantity
- 112 systems
- Distribution
- Worldwide Distribution --- including USA and countries of Belgium, Bulgaria, Canada, China, Denmark, France, Germany, Italy, Korea, Romania, Russia, Saudi Arabia, Singapore, Spain, Sweden, and United Kingdom.
- Lot, serial or model codes
- Serial numbers SG099, SG145, SG162, SG205, SG294, SG295, SG298, SG299, SG300, SG301, SG302, SG303, SG304, SG305, SG306, SG307, SG308, SG309, SG310, SG311, SG312, SG313, SG314, SG315, SG316, SG317, SG318, SG319, SG320, SG321, SG322, SG323, SG324, SG325, SG326, SG327, SG328, SG329, SG330, SG331, SG332, SG333, SG334, SG35, SG336, SG337, SG39, SG340, SG341, SG342, SG323, SG344, SG345, SG346, SG347, SG348, SG349, SG350, SG351, SG352, SG353, SG355, SG356, SG357, SG358, SG359, SG360, SG361, SG362, SG363, SG364, SG365, SG367, SG368, SG369, SG370, SG371, SG373, SG374, SG375, SG376, SG379, SG380, SG381, SG382, SG383, SG384, SG385, SG386, SG387, SG388, SG390, SG393, SG394, SG395, SG396, SG397, SG399, SG400, SG401, SG402, SG403, SG404, SG405, SG406, SG407, SG408, SG409, SG410, SG411, SG412, and SG413
- 510(k) numbers
- K050369, K060322
- PMA numbers
- not on file
- Product code
- NAY
- Firm FEI
- 3001675293
- Firm city
- Sunnyvale
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28