Philips Electronics North America Corporation: Pinnacle3 Software Version 9.0, 9.2, 9.4 and 9.6, Model Numbers 453560446041, 459800091001, 459800220161, 459800232931, 459800235871, 459800338451. UPDATED: Pinnacle3 Software Version 8.0h, 8.0k, 8.0m, 8.0n.

Recall number
Z-2200-2015
Recalling firm
Philips Electronics North America Corporation
Product
Pinnacle3 Software Version 9.0, 9.2, 9.4 and 9.6, Model Numbers 453560446041, 459800091001, 459800220161, 459800232931, 459800235871, 459800338451. UPDATED: Pinnacle3 Software Version 8.0h, 8.0k, 8.0m, 8.0n.
Status
Terminated
Initiated
2014-07-31
Posted
2015-07-22
Terminated
2019-05-08
Reason
Philips, Pinnacle Radiation Treatment Planning System version 8.0h, 8.0k, 8.0m, 8.0n, 9 0, 9 2 9 4, 9 6 is being recalled because the dose may be inconsistent with the density of a density-overridden ROl.
Root cause
Software design
Action
Consignees were sent on 7/31/2014 a Philips "Urgent Field Safety Notice" dated 2014-July 20. The letter described the problem, product involved in the recall and the action that should be taken by the customers. For questions contact Customer Care Solutions Center at 1-800-722-9377. UPDATE: A second Philips letter dated 2 July 2015 was sent in August 6, 2015. This letter updated the recall to incorporate the additional products being recalled.
Quantity
1642
Distribution
Worldwide Distribution-US: All states in the US including DC and PR. OUS: Australia, Austria, Bangladesh, Belgium, Brazil, China, Denmark, France, Germany, Hong Kong, India, Indonesia, Ireland, Italy, Japan, Malaysia, Mexico, Netherlands,New Zealand, Oman, Philippines, Singapore, South Africa, South Korea, Spain, Switzerland, Taiwan, Thailand, United Kingdom, Viet Nam. UPDATED: Canada, Bulgaria, Egypt, Hungary, Jordan, Luxembourg, Poland, Reunion, Romania, Russia, Sauth Arabia, Turkey.
Lot, serial or model codes
453560446041, 459800091001, 459800220161, 459800232931, 459800235871, 459800338451  UPDATED:  453560446051 [A]  459800096361 [Al 453560446061 IA]  459800096541 [A] 453560446091 [A]  459800096741 [A] 453560446101 [A]  459800096921 [A] 453560446111 [A]  459800097111 [A] 453560446131 [A]  459800097291 [A] 453560446141 [A]  459800097471 [A] 453560446161 [A]  459800097651 [B] 453560446181 [A]  459800097831 [B] 453560446201 [A]  459800098171 [B] 459800089221 [A]  459800098351 [A] 459800091451 [A]  459800109261 [A] 459800138101 [BJ 459800189431 [A) 459800220131 [B] 459800239171 [A] 459800233271 [8] 459800233281 [A] 459800233291 [B] 459800233301 [A] 459800233311 [8] 459800233321 [B] 459800233331 [B] 459800233341 [A] 459800233351 [A] 459800233361 [A] 459800234101 [A] 459800234111 [A] 459800234121 [A] 459800318391 [Al 459800405751 [A] 459800417961 [A] 459800249651 [8] 459800249721 [A] 459800249851 [A] 459800249931 [A] 459800250001 [Al 459800250081 [A] 459800250151 [A] 459800250221 [A] 459800250291 [A] 459800250361 [A] 459800250431 [A] 459800250501 [A] 459800250571 [A] 459800338961 [B] 459800340181 [A] 459800340201 [A] 459800341171 [A] 459800341191 [A] 459800341481 [A] 459800341501 [A] 459800345001 [A] 459800345021 [A] 459800345041 [A] 459800345061 [A] 459800345081 [A] 459800345101 [A] 459800454251 [A] 459800455341 [A] 459800493591 [A] 459800494011 [A] 459800554831 [A] 459800554881 [A] 453560435761 453560434881 453560454791 453560461331 453560461341
510(k) numbers
K041577
PMA numbers
not on file
Product code
LNH
Firm FEI
1218950
Firm city
Andover
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
3e6df46828c1
Size
288213574 bytes
Published
2026-10-06
Read
2026-10-07T03:20:25.099Z
Source file
Source file retained · verify
Refresh due
2026-10-14

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 3e6df46828c1, published 2026-10-06

Search this source