Medtronic Sofamor Danek USA Inc: T8 REMOVAL HEAD, REF 8910641, QTY: 1 EA, Product Usage: The non-sterile T8 Removal Diriver is used with the Spire and Spire Z Spinal System in conjunction with the set screwdriver to remove the implant during removal surgery. The set screw driver is attached to the end of the removal head and used to disengage the remaining portion of the set screw within the device.
- Recall number
- Z-2200-2013
- Recalling firm
- Medtronic Sofamor Danek USA Inc
- Product
- T8 REMOVAL HEAD, REF 8910641, QTY: 1 EA, Product Usage: The non-sterile T8 Removal Diriver is used with the Spire and Spire Z Spinal System in conjunction with the set screwdriver to remove the implant during removal surgery. The set screw driver is attached to the end of the removal head and used to disengage the remaining portion of the set screw within the device.
- Status
- Terminated
- Initiated
- 2013-08-12
- Posted
- 2013-09-14
- Terminated
- 2013-12-27
- Reason
- One lot of the T8 Removal Head had incorrect hardness specification, which could result in the T8 Removal head failing during removal of the CD Horizon Spire and Spire Z Spinal System implant. There is a potential for tissue damage to occur and/or minor surgical delay to remove the remaining fragments of the tip.
- Root cause
- Nonconforming Material/Component
- Action
- Medtronic sent an Urgent Voluntary Market Removal Recall letter dated August 9, 2013 with Recall Questionnaires and Confirmation forms included to Risk Managers and Sales Representatives. The letter identified the affected product, problem and action to be taken. Customers were instructed to remove the affected product, discontinue use and contact your Medtronic Sales Representative to coordinate the return and replenishment of affected product. For questions call your Medtronic Sales Representative or Global Quality Department at 800-876-3133.
- Quantity
- 9 units
- Distribution
- USA Nationwide Distribution in the state of GA, AK, WA, and CA.
- Lot, serial or model codes
- RS12K008
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- HXX
- Firm FEI
- 1000477302
- Firm city
- Memphis
- Firm state
- TN
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- 3e6df46828c1
- Size
- 288213574 bytes
- Published
- 2026-10-06
- Read
- 2026-10-07T03:20:25.099Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-14
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 3e6df46828c1, published 2026-10-06