Skip to the content

FDA Medical Device Recalls (openFDA)

Utah Medical Products, Inc: Intran Plus, Disposable Intrauterine Pressure Catheter System, Part numbers IUP-400, IUP-400S, IUP-450, IUP-450S, IUP-500, IUP-550, IUP-600, IUP-600S, IUP-650, IUP-650S, IUP-700, IUP-750, Sterile, Single Use, Utah Medical Products Inc., 7043 South 300 West, Midvale, Utah 84047.

Recall number
Z-2200-2008
Recalling firm
Utah Medical Products, Inc
Product
Intran Plus, Disposable Intrauterine Pressure Catheter System, Part numbers IUP-400, IUP-400S, IUP-450, IUP-450S, IUP-500, IUP-550, IUP-600, IUP-600S, IUP-650, IUP-650S, IUP-700, IUP-750, Sterile, Single Use, Utah Medical Products Inc., 7043 South 300 West, Midvale, Utah 84047.
Status
Terminated
Initiated
2008-06-16
Posted
2008-09-11
Terminated
2011-06-23
Reason
Possibility of compromised package sterility prior to the expiration date.
Root cause
Component change control
Action
All consignees were notified by a faxed Recall Notice letter beginning on 6/16/2008. Customers were asked to return all unused product for immediate replacement with units that have a protective barrier within the tray packaging. The firm cautioned customers to visually inspect packaging before use. Follow-up phone calls will be made the following week.
Quantity
300,000 units.
Distribution
Worldwide Distribution --- USA including states of CA, KS, MO, NC, NV, TX, VA, and WA and country of Japan.
Lot, serial or model codes
Lot numbers less than 1072755. Also 1080086, 1080087, 1080088, 1080089, 1080091, 1080093, 1080095, 1080245, 1080246, 1080247, 1080248, 1080249, 1080250, 1080266, 1080296, 1080297, 1080298, 1080299, 1080301, 1080302, 1080393, 1080394, 1080395, 1080432, 1080610, 1080669, and 1080721.
510(k) numbers
["K905563","K955443","K961424"]
PMA numbers
not on file
Product code
HFN
Firm FEI
1718873
Firm city
Midvale
Firm state
UT
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register