Stryker Corporation: Thermedx FluidSmart Urology Tube Set REF LL0006 UDI-DI code: 00860936000365 The FluidSmart Tube Sets are intended to be used with the FluidSmart Fluid Management System, which is intended for irrigation and fluid warming in laparoscopic procedures, and distention, fluid warming, and volume/deficit measurements in endoscopic procedures within gynecology, urology, and orthopedic disciplines.

Recall number
Z-2166-2026
Recalling firm
Stryker Corporation
Product
Thermedx FluidSmart Urology Tube Set REF LL0006 UDI-DI code: 00860936000365 The FluidSmart Tube Sets are intended to be used with the FluidSmart Fluid Management System, which is intended for irrigation and fluid warming in laparoscopic procedures, and distention, fluid warming, and volume/deficit measurements in endoscopic procedures within gynecology, urology, and orthopedic disciplines.
Status
Open, Classified
Initiated
2026-04-09
Posted
2026-05-12
Terminated
not on file
Reason
Due to nonconforming products being inadvertently distributed.
Root cause
Process change control
Action
On 04/09/2026, the firm sent via USPS Certified Mail an "URGENT: MEDICAL DEVICE RECALL" Letter to customers that Stryker identified that certain lots of FluidSmart tube sets were manufactured with an incorrect pressure relief valve specification. This defect may prevent the pressure relief valve from activating at the intended pressure in the event of a capital device over pressure failure. One lot (Lot B000576601, quantity 10) impacted by this nonconformance was inadvertently shipped to a customer. Customers are instructed to: 1. Inform individuals within their organization who need to be aware of this action. 2. Check internal inventory to locate the product listed in the product affected grid. 3. If affected product is found, stop use immediately, segregate the product and call Stryker Customer Service at 1-800-624-4422 (Option 3) or email endocustomersupport@stryker.com to arrange for product return and issuance of credit or replacement (upon availability). a. Remove the affected products, quarantine, and discontinue use of the identified product. b. Please complete Business Reply Form on Attachment A on Page 2 and return it to EndoRecall@stryker.com. 4. If affected product is NOT found: a. Please complete Business Reply Form on Attachment A on Page 2 and return it to EndoRecall@stryker.com. 5. Maintain awareness of this communication internally until all required actions have been completed within your facility. Response is required. For questions - EndoRecall@stryker.com or 1-800-624-4422 (Option 3).
Quantity
10 tube sets
Distribution
U.S. Nationwide distribution in the state of GA.
Lot, serial or model codes
Catalog Number: LL0006 UDI-DI code: 00860936000365 Lot Number: B000576601
510(k) numbers
K172048
PMA numbers
not on file
Product code
HIG
Firm FEI
2936485
Firm city
San Jose
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register