INNOVATIVE TOMOGRAPHY PRODUCT GMBH: Coaxial Interventional Needle, ITP innotom.com, CE 0297, STERILE EO for the following article numbers: Article no. KIM 22/05, Length 50 mm, diameter 22G/0,7 mm; Article no. KIM 22/07,Length 75 mm, diameter 22 G (0.7 mm); Article no. KIM 22/10, Length 100 mm, diameter 22 G (0.7 mm); Article no. KIM 20/15, Length 150 mm, diameter 20 G (0.9 mm); Article no. KIM 18/10, Length 100 mm, diameter 18 G (1.25 mm); Article no. KIM 18/10T, Length 100 mm, diameter 18 G (1.25 mm); Article no. KIM 18/15, Length 150 mm, diameter 18 G (1.25 mm); Article no. KIM 18/15T, Length 150 mm, diameter 18 G (1.25 mm); Article No. KIM 18/20, Length 200 mm, diameter 18G/1.27mm, Article no. KIM 13/04T, Length 44 mm, diameter 13 G (2.40 mm); Article no. KIM 13/09T, Length 94 mm , diameter 13 G (2.40 mm); Article no. KIM 13/14T, Length 144 mm, diameter 13 G (2.40 mm); Article no. KIM 15/04T, Length 44 mm, diameter 15 G (1.95 mm); Article no. KIM 15/09T, Length 94 mm, diameter 15 G (1.95 mm); Article no. KIM 15/14T, Length 144 mm, diameter 15 G (1.95 mm); Article no. KIM 16/04T,Length 44 mm, diameter 16 G (1.60 mm); Article no. KIM 16/09T, Length 94 mm, diameter 16 G (1.60 mm); Article no. KIM 16/14T, Length 144 mm, diameter 16 G (1.60 mm); Article no. KIM 14/20T, Length 200 mm, Diameter 14 G (2.1 mm)

Recall number
Z-2162-2021
Recalling firm
INNOVATIVE TOMOGRAPHY PRODUCT GMBH
Product
Coaxial Interventional Needle, ITP innotom.com, CE 0297, STERILE EO for the following article numbers: Article no. KIM 22/05, Length 50 mm, diameter 22G/0,7 mm; Article no. KIM 22/07,Length 75 mm, diameter 22 G (0.7 mm); Article no. KIM 22/10, Length 100 mm, diameter 22 G (0.7 mm); Article no. KIM 20/15, Length 150 mm, diameter 20 G (0.9 mm); Article no. KIM 18/10, Length 100 mm, diameter 18 G (1.25 mm); Article no. KIM 18/10T, Length 100 mm, diameter 18 G (1.25 mm); Article no. KIM 18/15, Length 150 mm, diameter 18 G (1.25 mm); Article no. KIM 18/15T, Length 150 mm, diameter 18 G (1.25 mm); Article No. KIM 18/20, Length 200 mm, diameter 18G/1.27mm, Article no. KIM 13/04T, Length 44 mm, diameter 13 G (2.40 mm); Article no. KIM 13/09T, Length 94 mm , diameter 13 G (2.40 mm); Article no. KIM 13/14T, Length 144 mm, diameter 13 G (2.40 mm); Article no. KIM 15/04T, Length 44 mm, diameter 15 G (1.95 mm); Article no. KIM 15/09T, Length 94 mm, diameter 15 G (1.95 mm); Article no. KIM 15/14T, Length 144 mm, diameter 15 G (1.95 mm); Article no. KIM 16/04T,Length 44 mm, diameter 16 G (1.60 mm); Article no. KIM 16/09T, Length 94 mm, diameter 16 G (1.60 mm); Article no. KIM 16/14T, Length 144 mm, diameter 16 G (1.60 mm); Article no. KIM 14/20T, Length 200 mm, Diameter 14 G (2.1 mm)
Status
Open, Classified
Initiated
2021-04-21
Posted
not on file
Terminated
not on file
Reason
Sterility Assurance; Coaxial Interventional Needle labeled as sterile may not have been adequately sterilized
Root cause
Vendor change control
Action
On April 20, 2021, Innovative Tomography Products GmbH issued a "urgent Safety Information Product Recall" notification to affected consignees via E-Mail. In addition to informing consignees about the recall products, the firm asked consignees mo to take the following actions: 1. We request that you immediately check your stock of the products in question with the LOT numbers indicated. Please block the corresponding products and quarantine them. 2. Please complete the attached response letter and return it to ITP via E-Mail to the following address: Address: ITP GmbH Rckrufunterlagen Universittsstrae 136 44799 Bochum Germany Contact person: Dominik Frmke Tel.: +49 234 54621942 Fax: +49 234 54622444 E-Mail: fraemke@innotom.com 3. Upon receipt of the documents, we will contact you to retrieve and exchange the products. Information sharing: 4. Please inform all persons within your company belonging to the circle of users about this urgent safety
Quantity
not on file
Distribution
US Nationwide distribution in the state of CA.
Lot, serial or model codes
Affected Lot Numbers:  KIM 22/05: LOT: 3416, 5016, 0717, 1417, 4117, 4217, 0418, 2818, 2119, 4619A, 3819, 0220A, 0220B, 0220C, 2420, 3820;  KIM 22/07: LOT: 4216, 0717, 1417, 4918, 2119, 4619A, 4319, 0220A, 0220B, 3820;  KIM 22/10: LOT: 3616, 5016, 0717, 1417, 0418, 1118, 2818, 2119, 2419, 4619A, 0220A, 0220B, 0220C, 4319, 1220, 1020, 2720, 3520;  KIM 22/15: LOT: 1417, 2118,4319;  KIM 20/10: LOT: 4318;  KIM 20/15: LOT: 4318;  KIM 18/10: LOT: 0717, 4318;  KIM 18/10T: LOT: 0217;  KIM 18/15: LOT: 0717, 4318, 0920;  KIM 18/15T: LOT: 4318;  KIM 18/20: LOT: 4318;  KIM 13/04T: LOT: 0817, 4318;  KIM 13/09T: LOT: 0817, 4318;  KIM 13/14T: LOT: 0817, 4318;  KIM 15/04T: LOT: 2818;  KIM 15/09T: LOT: 2818;  KIM 15/14T: LOT: 2818;  KIM 16/04T: LOT: 2818, 4820;  KIM 16/09T: LOT: 2818, 4820, 5020;  KIM 16/14T: LOT: 2818, 4820;  KIM 14/20T: LOT: 2419;
510(k) numbers
not on file
PMA numbers
not on file
Product code
GAA
Firm FEI
3014408997
Firm city
Bochum
Firm state
not on file
Firm country
Germany

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register