Skip to the content

FDA Medical Device Recalls (openFDA)

Microbiologics Inc: KWIK-STIK and LYFO DISK, quality control kit for culture media Catalog # 0805K and 0805L

Recall number
Z-2124-2025
Recalling firm
Microbiologics Inc
Product
KWIK-STIK and LYFO DISK, quality control kit for culture media Catalog # 0805K and 0805L
Status
Open, Classified
Initiated
2025-06-17
Posted
2025-07-17
Terminated
not on file
Reason
Affected products contain Staphylococcus epidermidis instead of the expected Brevundimonas diminuta. This may cause the user's quality control to fail and delayed diagnosis for the patient.
Root cause
Under Investigation by firm
Action
On June 17. 2025, the firm began notifying affected customers via "Urgent Medical Device Recall" letters delivered through email. Customers were instructed to review lab procedures to understand how the information may affect usage; use or discard product depending on lab procedures. Complete and return the response form to recall@microbiologics.com. If product was further distributed or transferred to other facilities or customers, they should also be notified. Customers may contact Microbiologics if a replacement kit is needed. If you have any questions or concerns, contact Recall Support team at 320.229.7080 or recall@microbiologics.com. Collect calls may be made.
Quantity
21 total
Distribution
US distribution to states of: Utah and Ohio; and OUS (International) to countries of: Italy, Netherlands, Argentina, Korea, Jordan, Spain, and Bangladesh.
Lot, serial or model codes
Lot #: 805-234-4, 805-234-5 UDI: (0805K) 30845357019531, (0805L) 10845357019544
510(k) numbers
["K861022"]
PMA numbers
not on file
Product code
JTR
Firm FEI
2150138
Firm city
Saint Cloud
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register