Skip to the content

FDA Medical Device Recalls (openFDA)

Philips Medical Systems (Cleveland) Inc: Ingenuity Core-Computed Tomography X-ray system Product Usage: The Ingenuity Core scanner is a whole body Computed Tomography X-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes.

Recall number
Z-2117-2016
Recalling firm
Philips Medical Systems (Cleveland) Inc
Product
Ingenuity Core-Computed Tomography X-ray system Product Usage: The Ingenuity Core scanner is a whole body Computed Tomography X-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes.
Status
Terminated
Initiated
2016-04-01
Posted
not on file
Terminated
2018-06-26
Reason
Philips Healthcare received reports from the field that certain Ingenuity Core systems running software version 3.5.4 exhibited intermittent swirl-like ring artifacts that may appear on reconstructed images. A patient rescan may be required if the images cannot be used for interpretation due to the swirl-like artifact.
Root cause
Software design (manufacturing process)
Action
Philips Medical Systems (Cleveland) Inc. sent an Urgent Field Safety Notice letter to affected customers , via certified letter with receipts returned for tracking purposes. The letter identified the affected product, problem and actions to be taken. Philips Healthcare is implementing software update version 3.5.5 to correct the problem. For questions contact your local Philips representative or local Philips Healthcare office.
Quantity
18
Distribution
US Nationwide Distribution in the states of CA, FL, ID, KS, MI, NY,OH,TX
Lot, serial or model codes
Ingenuity Core,System Code #728321, Serial numbers: 310101, 310102, 310103, 310104, 310105, 310106, 310108, 310109, 310112, 310115, 310137, 333006, 333008, 333011, 333012, 333013, 333016, 333017, 333023
510(k) numbers
K033326
PMA numbers
not on file
Product code
JAK
Firm FEI
1525965
Firm city
Cleveland
Firm state
OH
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
3e6df46828c1
Size
288213574 bytes
Published
2026-10-06
Read
2026-10-07T03:20:25.099Z
Source file
Source file retained · verify
Refresh due
2026-10-14

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 3e6df46828c1, published 2026-10-06

Search this source