FDA Medical Device Recalls (openFDA)
NeuroSync, Inc.: NeuroSync EYE-SYNC containing Pico Neo 2 Headsets. Each unit consists of a Pico Neo 2 headset with embedded sensors, a paired Android tablet, and proprietary software, packaged in a portable field case. Eye-tracking headset system is intended for recording, viewing, and analyzing eye movements in support of identifying visual tracking impairment in human subjects. The EYE-SYNC is intended to record, measure, and analyze eye movements as an aid in the diagnosis of concussion, also known as mild traumatic brain injury (mTBI).
- Recall number
- Z-2114-2025
- Recalling firm
- NeuroSync, Inc.
- Product
- NeuroSync EYE-SYNC containing Pico Neo 2 Headsets. Each unit consists of a Pico Neo 2 headset with embedded sensors, a paired Android tablet, and proprietary software, packaged in a portable field case. Eye-tracking headset system is intended for recording, viewing, and analyzing eye movements in support of identifying visual tracking impairment in human subjects. The EYE-SYNC is intended to record, measure, and analyze eye movements as an aid in the diagnosis of concussion, also known as mild traumatic brain injury (mTBI).
- Status
- Open, Classified
- Initiated
- 2023-09-19
- Posted
- 2025-07-11
- Terminated
- not on file
- Reason
- Eye syncing issues causing the device to be inoperable. Unable to be repaired due to discontinuation of support for the Pico 2 platform, which was no longer in production or serviceable as of August 31, 2023.
- Root cause
- Device Design
- Action
- NeuroSync notified affected customers beginning on 09/19/2024 via direct email and telephone calls. Customers were instructed to discontinue use of the affected Pico 2 device, return or decommission the unit, and transition to the new replacement system provided (Pico 3 platform).
- Quantity
- 27 units
- Distribution
- United States Nationwide distribution in the states of New York (NY), Texas (TX), and Georgia (GA).
- Lot, serial or model codes
- Model No X02. UDI-DI: 00868096000300. Serial Numbers: PA7D50NGF4020311G (Lot P-10114-13), PA7D50NGF4020191G (Lot P-10114-13), PA7D40NGEA080237G (Lot P-10114-12) Software Revisions: Software version uniform across all affected devices
- 510(k) numbers
- ["K202927"]
- PMA numbers
- not on file
- Product code
- QEA
- Firm FEI
- 3014133165
- Firm city
- Holliston
- Firm state
- MA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as