Skip to the content

FDA Medical Device Recalls (openFDA)

International Rehabilitative Sciences, Inc.: The "RS-FBG Full Back Conductive Garment" with electrode attachments for use with interferential and muscle stimulation with a Transcutaneous Electrical Nerve Stimulation device (TENS) and/or muscle stimulator. The RS-FBG is a prescription device for use with RS Medical Stimulators to facilitate electrode placement and maintenance of electrode positioning for stimulation treatments. The device is sold individually in several variations for size. It is fitted to the patient and an RS Medical employee gives the patient instructions on how to utilize the device with proper electrode placement in conjunction with the respective TENS/muscle stimulator.

Recall number
Z-2093-2010
Recalling firm
International Rehabilitative Sciences, Inc.
Product
The "RS-FBG Full Back Conductive Garment" with electrode attachments for use with interferential and muscle stimulation with a Transcutaneous Electrical Nerve Stimulation device (TENS) and/or muscle stimulator. The RS-FBG is a prescription device for use with RS Medical Stimulators to facilitate electrode placement and maintenance of electrode positioning for stimulation treatments. The device is sold individually in several variations for size. It is fitted to the patient and an RS Medical employee gives the patient instructions on how to utilize the device with proper electrode placement in conjunction with the respective TENS/muscle stimulator.
Status
Terminated
Initiated
2010-05-17
Posted
not on file
Terminated
2012-08-15
Reason
Device distributed with labeling and IFU outside of their 510(k) clearance.
Root cause
Other
Action
The firm, RS Medical, mailed "URGENT NOTICE OF VOLUNTARY CORRECTIVE ACTION" letters dated May 14, 2010, to customers on May 17, 2010 and letters to physicians/consignees on May 19, 2010. The letters included an updated RS-FBG Operation Manual and stated what the update covered; i.e. that the device is indicated for use on the "middle and upper back". The customers were ask to please read your new manual and replace your original RS-FBG Full Back Conductive Garment Operation Manual with this updated version and discard your old RS-FBG Garment Operation Manual. RS Medical stated that your RS-FBG Garment is not affected by this Corrective Action, only the Operation Manual. RS Medical informed the customers that you may continue to use the RS-FBG Full Back Garment with your RS Medical Stimulator as prescribed by your healthcare provider. If you have any questions or concerns, please call our Toll Free Customer Service number at 1-866-417-1293, email us at RSFBG@rsmedical.com or visit our special web page at www.rsmedical.com/rsfbginfo.
Quantity
108134 units, 81,939 subject to recall
Distribution
Nationwide distribution.
Lot, serial or model codes
************All serial numbers from "GA133809" and lower. Code Breakdown: "GA" specific for Full Back Garments; "133809" is a six digit sequential number as devices are manufactured.
510(k) numbers
K090951
PMA numbers
not on file
Product code
GXY
Firm FEI
1000117825
Firm city
Vancouver
Firm state
WA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
3e6df46828c1
Size
288213574 bytes
Published
2026-10-06
Read
2026-10-07T03:20:25.099Z
Source file
Source file retained · verify
Refresh due
2026-10-14

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 3e6df46828c1, published 2026-10-06

Search this source