Skip to the content

FDA Medical Device Recalls (openFDA)

Mckesson Medical-Surgical Inc. Corporate Office: ANCILLARY ADULT 140-DOSE CONVENIENCE KIT, ITEM #1189482, for the administration of the Moderna COVID-19 vaccine convenience kit for the administration of the COVID-19 vaccine

Recall number
Z-2081-2021
Recalling firm
Mckesson Medical-Surgical Inc. Corporate Office
Product
ANCILLARY ADULT 140-DOSE CONVENIENCE KIT, ITEM #1189482, for the administration of the Moderna COVID-19 vaccine convenience kit for the administration of the COVID-19 vaccine
Status
Terminated
Initiated
2021-06-02
Posted
not on file
Terminated
2022-04-28
Reason
Two unsuitable 3mL syringes with conventional needles for vaccine administration included in some Moderna 140 Ancillary Convenience Kits.
Root cause
Process change control
Action
McKesson issued an "URGENT MEDICAL DEVICE SAFETY ALERT" on 06/02/2021 by email to its consignees. The notice requested that the consignees do the following: -examine their inventory immediately to determine if you have any quantities of the product, -Immediately stop using these syringes with 3mI syringes with conventional needles for the administration of the Moderna COVID-19 vaccine. -As these 3mI syringes with conventional needles can be utilized for a variety of other uses, such as medication mixing in healthcare settings, you may continue to use these 3mI syringes with conventional needles in your healthcare facility for other appropriate uses. -Please fill out and return the Medical Device Alert Response Form attached. The firm would be shipping replacement safety syringes. For questions about this notification, please call 877-546-5661.
Quantity
53,388 kits
Distribution
US Nationwide Distribution
Lot, serial or model codes
All convenience kit lot numbers distributed 5/4/2021 - 5/24/2021, containing syringes as follows: a) SYRINGE/NDL, 3CC 25GX1 1/2" BD, Lot numbers: 9252960, 0296288, 9252961 b) SYRINGE/NDL, 3CC 23GX1 1/2" BD, Lot Numbers: 0324763, 0324761, 0324766, 0324770, 1015383, 1020431
510(k) numbers
not on file
PMA numbers
not on file
Product code
QMT
Firm FEI
3004386693
Firm city
Richmond
Firm state
VA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register