FDA Medical Device Recalls (openFDA)
Etac Supply Center Ab: Etac Supporter toilet armrest Etac Supporter toilet armrests are foldable and have comfortable rounded grips for the hands. Supporter is simple to mount without the need for any tools. It fits both floor and wall mounted toilets. If space around the toilet is limited either armrest can easily be detached. Supporter includes seat ring and lid.
- Recall number
- Z-2073-2015
- Recalling firm
- Etac Supply Center Ab
- Product
- Etac Supporter toilet armrest Etac Supporter toilet armrests are foldable and have comfortable rounded grips for the hands. Supporter is simple to mount without the need for any tools. It fits both floor and wall mounted toilets. If space around the toilet is limited either armrest can easily be detached. Supporter includes seat ring and lid.
- Status
- Terminated
- Initiated
- 2015-06-15
- Posted
- 2015-07-15
- Terminated
- 2017-07-14
- Reason
- The Supporter toilet armrest has two Fast Nuts fixing the product to the toilet porcelain. The firm has reports of the two Fast Nuts coming loose and causing the Supporter to become unsteady.
- Root cause
- Nonconforming Material/Component
- Action
- Etac sent an Urgent Notice of Field Correction beginning June 15, 2015, to all affected consignees. The letter directed the consignees to do an in field correction with replacement Fast Nuts. Defined replacement instructions were included with the replacement nuts. Customers with questions were instructed to contact their Etac distributor or the Etac coordinator, Snug Seal Inc. at 1-800-336-7683. For questions regarding this recall call 1-800-367-6844.
- Quantity
- 25 units
- Distribution
- Nationwide Distribution including GA, CT and PA.
- Lot, serial or model codes
- Serial numbers higher than 224017
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- ILS
- Firm FEI
- 3008986539
- Firm city
- Anderstorp
- Firm state
- not on file
- Firm country
- Sweden
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28