Philips North America, LLC: HeartStart MRx Processor Board PCA Replacement Kits 453563478461, a component of the Philips HeartStart MRx Monitorl/Defibrillator - Product Usage: is intended for use in hospital and prehospital settings by qualified medical personnel trained in the operation of the device and qualified by training in basic life support, advanced cardiac life support or defibrillation.

Recall number
Z-2064-2020
Recalling firm
Philips North America, LLC
Product
HeartStart MRx Processor Board PCA Replacement Kits 453563478461, a component of the Philips HeartStart MRx Monitorl/Defibrillator - Product Usage: is intended for use in hospital and prehospital settings by qualified medical personnel trained in the operation of the device and qualified by training in basic life support, advanced cardiac life support or defibrillation.
Status
Terminated
Initiated
2020-04-10
Posted
not on file
Terminated
2021-02-10
Reason
Damaged ESD bags used for storing components in Processor PCA Replacement Kits
Root cause
Process control
Action
Philips Healthcare ssued Medical Device Correction letter dated April 10, 2020 via USP stating reason for correction, health risk and action to take: If the Processor Board PCA Replacement Kit has not yet been installed in a Heart5tart MRx Monitor/Defibrillator, quarantine the kit until Philips can replace it. If the PCA has been installed in a HeartStart MRx Monitor/Defibrillator, continue to monitor the status of the device as recommended in the Instructions for Use, including performing Operational Checks routinely as recommended in the Instructions for Use. Verify your MRx does not show a red X nor is it emitting a periodic audio chirp as this could be an indicator a failure has been detected. When the device is turned on, the device will display INOP messages for any failures detected. If a failure has been detected, take the device out of use and contact your local Philips representative. If an affected Processor Board PCA Replacement Kit 453563478461 is identified, please contact your local Phillips representative. Once the customer has contacted their local Philips representative with their HeartStart MRx Monitor/Defibrillator serial number and also confirmed the affected Processor Board PCA serial number, Philips will provide a replacement Processor Board PCA 453563478461 Replacement Kit free of charge. Further information or support concerning this issue, contact your local Philips representative or call us at 1-800-722-9377.
Quantity
8 units (2 US and 6 OUS)
Distribution
Worldwide distribution - US Nationwide distribution including in the states of OK, NY and the countries of Libya, Netherlands, Canada, South Africa, United Kingdom.
Lot, serial or model codes
Serial Numbers: CO1-1643IDOTDB CO1-1639100N3T C01-16391D0NB2, CO1-1639100N43 C01-1639100ND0 C01-16391D0N6N, C01-16391D0N4N C01-1&39IDON1J
510(k) numbers
["K063375"]
PMA numbers
not on file
Product code
MKJ
Firm FEI
1218950
Firm city
Andover
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register