Centinel Spine, Inc.: Prodisc C SK U.S. IMPLANT EXTRA LARGE 6MM. Model Number: PDSXL6. Total cervical disc replacement.
- Recall number
- Z-2051-2026
- Recalling firm
- Centinel Spine, Inc.
- Product
- Prodisc C SK U.S. IMPLANT EXTRA LARGE 6MM. Model Number: PDSXL6. Total cervical disc replacement.
- Status
- Open, Classified
- Initiated
- 2026-03-20
- Posted
- 2026-05-01
- Terminated
- not on file
- Reason
- Product labeling mix up. The prodisc C SK U.S. Implant Extra Large 6mm product was labeled as a 5mm and prodisc C SK U.S. Implant Extra Large 5mm product was labeled as a 6mm.
- Root cause
- Labeling mix-ups
- Action
- Centinel Spine notified distributors on about 03/20/2026 via telephone calls and emailed letter. Surgeons were sent letters via FedEx on 03/24/2026 and by email on 03/25/2026 & 03/26/2026. Distributors were instructed to identify any affected units on hand, provide photographs documenting such, and arrange for the return of all affected units. Additionally, distributors have been instructed to diligently check prodisc C SK inventory prior to any scheduled surgery to ensure that the recalled lot numbers are not available for implantation. This includes both consignment and loaner inventory. Acknowledgment of the letter has been requested with return envelopes provided. For surgeons, they were instructed to monitor any patient(s) that have received the recalled device for any unexpected device-related symptoms throughout the regularly scheduled check-in procedures. Centinel Spine intends to remove from the market all remaining implants within the lots effected by this recall and issue a replacement product with corrected labeling for any remaining products that are subject to this recall.
- Quantity
- 30 units
- Distribution
- US Nationwide distribution in the states of AZ, CA, GA, LA, MO, NY, TN, TX.
- Lot, serial or model codes
- Model Number: PDSXL6; UDI-DI: 00843193113986; Lot Number: 2026-0027
- 510(k) numbers
- not on file
- PMA numbers
- P070001, P070001S019, P070001S025
- Product code
- MJO
- Firm FEI
- 3007494564
- Firm city
- West Chester
- Firm state
- PA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28