Philips North America Llc: DigitalDiagnost C90, Model No. 712034 and 712035
- Recall number
- Z-2042-2023
- Recalling firm
- Philips North America Llc
- Product
- DigitalDiagnost C90, Model No. 712034 and 712035
- Status
- Open, Classified
- Initiated
- 2023-05-19
- Posted
- 2023-06-30
- Terminated
- not on file
- Reason
- There is a software login in issue that may prevent the user from logging in.
- Root cause
- Software change control
- Action
- Consignees in the United States were notified via Urgent Medical Device Correction Letter that was distributed on May 26, 2023. US consignees were initially notified via email. Follow-up notifications via FedEx will follow. Philips Market Organizations are responsible for distributing letters outside of the US in accordance with local requirements. A Philips Field Service Engineer will visit impacted customer sites to perform a software upgrade to prevent the issue from recurring. Philips has requested customers perform the following: a) To prevent the system from becoming inoperable, do not reboot or log out of your system until Philips has implemented the software correction; b) Circulate this notice to all users of the device so that they are aware of the issue; and c) Complete and return the response form to Philips promptly and no later than 30 days from receipt.
- Quantity
- 58
- Distribution
- Distribution in US: AZ, GA, IN, IA, MN, NE, NH, OH, TX OUS distribution to Australia, Chile, China, Malta, Namibia, New Zealand, Oman, Panama, Qatar, Suriname, United Kingdom
- Lot, serial or model codes
- 1) UDI-DI (01)00884838090699, Model 712034, serial numbers: 10000074, 10000083, 10000163, 10000188, 10001057, 10001071, 10001072, 10001079, 10001089, 10001106, 10001133, 10001136, 10001137, 10001142, 10001146, 10001169, 10001180, 10001181, 10001186, 10001203, 10001211, 10001229, 10001234, 10001276, 19861179, 19861365, 21860026, 21860084, 21861001, 21861233, 21861386, 21861415, 21861416, 21861493, 21861528, 22860017, 22860114, 22860115, 22860448, 22860454, SN19000006, SN20861092, SN20861107, SN20861108, SN20861115, SN20861125, SN20861126, and SN20861128 2) UDI-DI (01)00884838090705, Model 712035, serial numbers: 10000034, 10000098, 10001051, 10001113, 10001130, 21860040, 22860052, SN20862021, SN20862022, and SN20862024
- 510(k) numbers
- ["K202564"]
- PMA numbers
- not on file
- Product code
- MQB
- Firm FEI
- 3016618143
- Firm city
- Cambridge
- Firm state
- MA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28