Philips North America Llc: DigitalDiagnost C90, Model No. 712034 and 712035

Recall number
Z-2042-2023
Recalling firm
Philips North America Llc
Product
DigitalDiagnost C90, Model No. 712034 and 712035
Status
Open, Classified
Initiated
2023-05-19
Posted
2023-06-30
Terminated
not on file
Reason
There is a software login in issue that may prevent the user from logging in.
Root cause
Software change control
Action
Consignees in the United States were notified via Urgent Medical Device Correction Letter that was distributed on May 26, 2023. US consignees were initially notified via email. Follow-up notifications via FedEx will follow. Philips Market Organizations are responsible for distributing letters outside of the US in accordance with local requirements. A Philips Field Service Engineer will visit impacted customer sites to perform a software upgrade to prevent the issue from recurring. Philips has requested customers perform the following: a) To prevent the system from becoming inoperable, do not reboot or log out of your system until Philips has implemented the software correction; b) Circulate this notice to all users of the device so that they are aware of the issue; and c) Complete and return the response form to Philips promptly and no later than 30 days from receipt.
Quantity
58
Distribution
Distribution in US: AZ, GA, IN, IA, MN, NE, NH, OH, TX OUS distribution to Australia, Chile, China, Malta, Namibia, New Zealand, Oman, Panama, Qatar, Suriname, United Kingdom
Lot, serial or model codes
1) UDI-DI (01)00884838090699, Model 712034, serial numbers: 10000074, 10000083, 10000163, 10000188, 10001057, 10001071, 10001072, 10001079, 10001089, 10001106, 10001133, 10001136, 10001137, 10001142, 10001146, 10001169, 10001180, 10001181, 10001186, 10001203, 10001211, 10001229, 10001234, 10001276, 19861179, 19861365, 21860026, 21860084, 21861001, 21861233, 21861386, 21861415, 21861416, 21861493, 21861528, 22860017, 22860114, 22860115, 22860448, 22860454, SN19000006, SN20861092, SN20861107, SN20861108, SN20861115, SN20861125, SN20861126, and SN20861128 2) UDI-DI (01)00884838090705, Model 712035, serial numbers: 10000034, 10000098, 10001051, 10001113, 10001130, 21860040, 22860052, SN20862021, SN20862022, and SN20862024
510(k) numbers
["K202564"]
PMA numbers
not on file
Product code
MQB
Firm FEI
3016618143
Firm city
Cambridge
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register