Skip to the content

FDA Medical Device Recalls (openFDA)

NordicNeuroLab AS: nordicTumorEx 1.0 nordicTumorEx is an image processing software package to be used by trained professionals including but not limited to physicians and medical technicians. The software runs on a standard off-the-shelf PC workstation and can be used to perform image viewing, processing and analysis of medical images. Data and images are acquired through DICOM compliant imaging devices and modalities. nordicTumorEx software enables processing and visualization of dynamic MR imaging datasets of the brain, showing properties of changes in contrast overtime. It generates various parametric images calculated from the image intensity variations and provides tools to extract and visualize such parametric properties within specified volumes of interest. Structural datasets providing anatomical information can be sued to evaluate the extent of a specified sub-volume, such as a tumor. Comparison of imaging studies performed at different study dates may also be performed to support the diagnostic process.

Recall number
Z-2042-2018
Recalling firm
NordicNeuroLab AS
Product
nordicTumorEx 1.0 nordicTumorEx is an image processing software package to be used by trained professionals including but not limited to physicians and medical technicians. The software runs on a standard off-the-shelf PC workstation and can be used to perform image viewing, processing and analysis of medical images. Data and images are acquired through DICOM compliant imaging devices and modalities. nordicTumorEx software enables processing and visualization of dynamic MR imaging datasets of the brain, showing properties of changes in contrast overtime. It generates various parametric images calculated from the image intensity variations and provides tools to extract and visualize such parametric properties within specified volumes of interest. Structural datasets providing anatomical information can be sued to evaluate the extent of a specified sub-volume, such as a tumor. Comparison of imaging studies performed at different study dates may also be performed to support the diagnostic process.
Status
Terminated
Initiated
2014-12-15
Posted
not on file
Terminated
2021-03-24
Reason
An error was discovered in the interpretation of certain DICOM header tags that may lead to incorrect orientation labeling, and thus and indirect left-right, up-down or anterior-posterior flipping of images.
Root cause
Software design
Action
The firm notified their consignees of the software problem by email on December 14, 2014. For further questions please call (262) 337-2909.
Quantity
16 licenses
Distribution
Worldwide Distribution - USA (nationwide) and to the countries of : SA, Russia, South Korea, UK, Japan, Australia, Sweden, Norway, Greece
Lot, serial or model codes
Version 1.0.4
510(k) numbers
["K123306"]
PMA numbers
not on file
Product code
LLZ
Firm FEI
3006738448
Firm city
Bergen
Firm state
not on file
Firm country
Norway

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register