Skip to the content

FDA Medical Device Recalls (openFDA)

GE Medical Systems Information Technologies Inc: GE HealthCare CRITIKON ONE-CUF blood pressure cuffs, Model numbers: 1. ONE-A1-1B, ONE-CUF, SMALL ADULT, 1 TB BAYONET, 17 - 25 CM, 20/BOX; 2. ONE-A1-2A, ONE-CUF, SMALL ADULT, DINACLICK 80369-5, 17 - 25 CM, 20/BOX; 3. ONE-A2-1B, ONE-CUF, ADULT, 1 TB BAYONET, 23 - 33 CM, 20/BOX; 4. ONE-A1-1B-L, ONE-CUF, ADULT LONG, 1 TB BAYONET, 23 - 33 CM, 20/BOX; 5. ONEA2-2A, ONE-CUF, ADULT, DINACLICK 80369-5, 23 - 33 CM, 20/BOX; 6. ONE-A2-2A-L, ONE-CUF, ADULT LONG, DINACLICK 80369-5, 23 - 33 CM, 20/BOX; 7. ONE-A3-1B, ONE-CUF, LARGE ADULT, 1 TB BAYONET, 31 - 40 CM, 20/BOX; 8. ONE-A3-2A, ONE-CUF, LARGE ADULT, DINACLICK 80369-5, 31 - 40 CM, 20/BOX; 9. ONE-P1-1B, ONE-CUF, INFANT, 1 TB BAYONET, 08 - 13 CM, 20/BOX; 10. ONE-P1-2A, ONE-CUF, INFANT, DINACLICK 80369-5, 08 - 13 CM, , 20/BOX; 11. ONE-P2-1B, ONE-CUF, CHILD, 1 TB BAYONET, 12 - 19 CM, 20/BOX; and 12. ONE-P2-2A, ONE-CUF, CHILD, DINACLICK 80369-5, 12 - 19 CM, 20/BOX.

Recall number
Z-2027-2024
Recalling firm
GE Medical Systems Information Technologies Inc
Product
GE HealthCare CRITIKON ONE-CUF blood pressure cuffs, Model numbers: 1. ONE-A1-1B, ONE-CUF, SMALL ADULT, 1 TB BAYONET, 17 - 25 CM, 20/BOX; 2. ONE-A1-2A, ONE-CUF, SMALL ADULT, DINACLICK 80369-5, 17 - 25 CM, 20/BOX; 3. ONE-A2-1B, ONE-CUF, ADULT, 1 TB BAYONET, 23 - 33 CM, 20/BOX; 4. ONE-A1-1B-L, ONE-CUF, ADULT LONG, 1 TB BAYONET, 23 - 33 CM, 20/BOX; 5. ONEA2-2A, ONE-CUF, ADULT, DINACLICK 80369-5, 23 - 33 CM, 20/BOX; 6. ONE-A2-2A-L, ONE-CUF, ADULT LONG, DINACLICK 80369-5, 23 - 33 CM, 20/BOX; 7. ONE-A3-1B, ONE-CUF, LARGE ADULT, 1 TB BAYONET, 31 - 40 CM, 20/BOX; 8. ONE-A3-2A, ONE-CUF, LARGE ADULT, DINACLICK 80369-5, 31 - 40 CM, 20/BOX; 9. ONE-P1-1B, ONE-CUF, INFANT, 1 TB BAYONET, 08 - 13 CM, 20/BOX; 10. ONE-P1-2A, ONE-CUF, INFANT, DINACLICK 80369-5, 08 - 13 CM, , 20/BOX; 11. ONE-P2-1B, ONE-CUF, CHILD, 1 TB BAYONET, 12 - 19 CM, 20/BOX; and 12. ONE-P2-2A, ONE-CUF, CHILD, DINACLICK 80369-5, 12 - 19 CM, 20/BOX.
Status
Open, Classified
Initiated
2024-04-09
Posted
2024-06-06
Terminated
not on file
Reason
Use of the blood pressure cuffs could result in inaccurate non-invasive blood pressure measurements.
Root cause
Under Investigation by firm
Action
Recall letters dated and issued on 4/9/2024 by the recalling firm explained the safety issue and actions to be taken by the customer/user. Those actions were to not use the affected products listed in the Affected Product Details section of the letter, use an alternate cuff for NIBP monitoring, and destroy all affected product per the customer facility procedures. A photograph of the product was embedded in the letter to show where to find the product name, model number, and GTIN number. The letter informs the consignee that GE HealthCare will replace the affected products with CRITIKON SOFT-CUF Blood Pressure Cuffs and a GE HealthCare representative will contact the consignee to arrange for this correction. An Acknowledgement Form was enclosed for completion and return.
Quantity
23,180/20-cuff boxes
Distribution
US Nationwide distribution in the states of AL, AR, AZ, CA, FL, GA, IA, IL, IN, LA, MD, MO, MS, NC, NJ, NM, OH, PA, TN, and TX.
Lot, serial or model codes
No serial numbers are on the blood pressure cuffs. GTIN NUMBERS: ONE-A1-1B - GTIN 00195278573803; ONE-A1-2A - GTIN 00195278576460; ONE-A2-1B - GTIN 00195278573797; ONE-A1-1B-L - GTIN 00195278573780; ONEA2-2A - GTIN 00195278576477; ONE-A2-2A-L GTIN 00195278576484; ONE-A3-1B - GTIN 00195278573810; ONE-A3-2A - GTIN 00195278576491; ONE-P1-1B - GTIN 00195278573834; ONE-P1-2A - GTIN 00195278576507; ONE-P2-1B - GTIN 00195278573827; and ONE-P2-2A - GTIN 00195278576514.
510(k) numbers
["K230145"]
PMA numbers
not on file
Product code
DXQ
Firm FEI
3012359877
Firm city
Milwaukee
Firm state
WI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register