Medtronic Navigation, Inc.-Littleton: O-arm O2 Imaging System-Mobile x-ray system designed for 2D fluoroscopic and 3D imaging for adult and pediatric patients. Product Number: BI70002000

Recall number
Z-2026-2024
Recalling firm
Medtronic Navigation, Inc.-Littleton
Product
O-arm O2 Imaging System-Mobile x-ray system designed for 2D fluoroscopic and 3D imaging for adult and pediatric patients. Product Number: BI70002000
Status
Open, Classified
Initiated
2024-04-26
Posted
2024-06-06
Terminated
not on file
Reason
Potential for an electrical component of the O-arm" O2 Imaging System to intermittently shut off,failure can cause the potential for surgical delay, additional imaging, cancellation and rescheduling of surgery, or completion of surgery without the benefit of image guidance.
Root cause
Device Design
Action
Medtronic issued Urgent Medical Device Correction Letter on 4/26/24 to US end-users hand delivered via Field Service representative. OUS communications were initiated with affected consignees on April 17, 2024. Letter states reason for recall, health risk and action to take: A Medtronic Field Service Engineer will contact your facility to schedule time to service your system within 90 days. Required Actions: 1. If your system is actively in use, you may continue to use it pending service by Medtronic. If your system exhibits the issue, it will automatically reset after beeping. If the frequency of the issue impacts your ability to ensure the system can be in place for procedures, an alternate imaging system should be considered. 2. If your system displays a placard identifying this issue, the placard must remain attached to the system. Your Medtronic Field Service Engineer will remove the placard upon completion of service. 3. Use the enclosed Customer Confirmation Form to confirm receipt of this notification letter. Send the completed Customer Confirmation Form to Medtronic via email at neuro.quality@medtronic.com. 4. This notice needs to be passed on to those who need to be aware within your organization or to any organization where the potentially affected devices have been transferred. Please maintain a copy of this notice in your records. Additional Information: If you have any questions related to this issue, please contact your Medtronic Field Service Engineer or Technical Services for help at 1-888-826-5603 (24 hours/ 7 days per week) or email at rs.navtechsupport@medtronic.com.
Quantity
20 units
Distribution
Worldwide - US Nationwide distribution in the states of AZ, CA, CO, CT, FL, MN, PR, TX, WI and the countries of Bangladesh, Canada, Costa Rica, India, Mexico, Panama, Philippines.
Lot, serial or model codes
GTIN: 00643169639683 00763000616441 00763000616465 00763000616526 00763000616533 00763000616564 00763000616571 Serial Numbers: C3578 C3421 C3606 C3645 C3641 C3589 C3604 C3537 C3551 C1595 C1437 C3469 C3635 C3626 C3389 C3647 C3407 C3713 C3607 C1629
510(k) numbers
K151000
PMA numbers
not on file
Product code
OWB
Firm FEI
3004785967
Firm city
Littleton
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register