Skip to the content

FDA Medical Device Recalls (openFDA)

Cardiogenesis Corporation: Cardiogenesis Corporation - Cardiogenesis Transmyocardial Laser. The device is indicated for use in the treatment of patients with angina (Canadian Cardiovascular Society Class 4) refractory to medical treatment and secondary to objectively demonstrated coronary artery atherosclerosis and with a region of the myocardium with reversible ischemia not amenable to direct coronary revascularization.

Recall number
Z-1993-2013
Recalling firm
Cardiogenesis Corporation
Product
Cardiogenesis Corporation - Cardiogenesis Transmyocardial Laser. The device is indicated for use in the treatment of patients with angina (Canadian Cardiovascular Society Class 4) refractory to medical treatment and secondary to objectively demonstrated coronary artery atherosclerosis and with a region of the myocardium with reversible ischemia not amenable to direct coronary revascularization.
Status
Terminated
Initiated
2011-05-01
Posted
2013-09-11
Terminated
2014-09-15
Reason
Following acquisition of Cardiogenesis, we began receiving complaints of sparks, burns, or smoke incidents related to the handpieces.
Root cause
Device Design
Action
Cardiogenesis is responsible for the correction of all noncompliant products. CDRH understands that the corrective action plan (CAP), involving 2,437 Sologrip III handpieces and 61 PEARL 5.0 handpieces, consists of the following: 1. Notification to purchasers of the handpiece removal and replacement procedures, 2. Replace existing hospital inventory of handpieces with a redesigned handpiece CDRH approves the CAP subject to the following conditions: 1. Notification of all distributors and purchasers will be successfully implemented, and 2. Corrections will be made at no cost to the purchasers by October 1, 2013. For further questions please call (949) 420-1800.
Quantity
2498
Distribution
US Distribution
Lot, serial or model codes
Product Code - HP-SG3 Product Code -HP-PRL5
510(k) numbers
not on file
PMA numbers
not on file
Product code
MNO
Firm FEI
3001451326
Firm city
Kennesaw
Firm state
GA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28