Ellex Medical Pty Ltd.: Integre Pro Yellow LP561, Integre Pro Yellow LP6RY, Integre Pro Scan Green LP6G, Integre Pro Green LP5532 Spare, Optics Bench V2, L2G/LP6G

Recall number
Z-1971-2023
Recalling firm
Ellex Medical Pty Ltd.
Product
Integre Pro Yellow LP561, Integre Pro Yellow LP6RY, Integre Pro Scan Green LP6G, Integre Pro Green LP5532 Spare, Optics Bench V2, L2G/LP6G
Status
Open, Classified
Initiated
2023-03-16
Posted
2023-06-27
Terminated
not on file
Reason
When using a Laser-Indirect-Ophthalmoscope (LIO) as the delivery device, a scattered laser beam inside the laser cavity system may unintentionally exit through the Slit Lamp output port (SDS) and get emitted from the Objective Lens when the laser is fired.
Root cause
Radiation Control for Health and Safety Act
Action
Ellex Safety Bulletin is the customer notification letter. Ellex will, without charge, remedy the defect or bring the product into compliance with each applicable Federal standard in accordance with a plan to be approved by the Secretary of Health and Human Services. For questions contact your local Ellex-Lumibird representative or email us on eservice@ellex.com.
Quantity
57 systems in total (6 units shipped to United States)
Distribution
US nationwide/ Worldwide Distribution
Lot, serial or model codes
Serial Nos., UDI IPY0374, (01) 0 9342395 00012 0 ISRY0242, (01) 0 9342395 00015 1 ISG0370, (01) 0 9342395 00024 3 IPG0539, (01) 0 9342395 00009 0 BG1190
510(k) numbers
["K081565","K142398"]
PMA numbers
not on file
Product code
HQF
Firm FEI
3002806902
Firm city
Lakes
Firm state
Mawson
Firm country
Australia

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register