FDA Medical Device Recalls (openFDA)
Ellex Medical Pty Ltd.: Integre Pro Yellow LP561, Integre Pro Yellow LP6RY, Integre Pro Scan Green LP6G, Integre Pro Green LP5532 Spare, Optics Bench V2, L2G/LP6G
- Recall number
- Z-1971-2023
- Recalling firm
- Ellex Medical Pty Ltd.
- Product
- Integre Pro Yellow LP561, Integre Pro Yellow LP6RY, Integre Pro Scan Green LP6G, Integre Pro Green LP5532 Spare, Optics Bench V2, L2G/LP6G
- Status
- Open, Classified
- Initiated
- 2023-03-16
- Posted
- 2023-06-27
- Terminated
- not on file
- Reason
- When using a Laser-Indirect-Ophthalmoscope (LIO) as the delivery device, a scattered laser beam inside the laser cavity system may unintentionally exit through the Slit Lamp output port (SDS) and get emitted from the Objective Lens when the laser is fired.
- Root cause
- Radiation Control for Health and Safety Act
- Action
- Ellex Safety Bulletin is the customer notification letter. Ellex will, without charge, remedy the defect or bring the product into compliance with each applicable Federal standard in accordance with a plan to be approved by the Secretary of Health and Human Services. For questions contact your local Ellex-Lumibird representative or email us on eservice@ellex.com.
- Quantity
- 57 systems in total (6 units shipped to United States)
- Distribution
- US nationwide/ Worldwide Distribution
- Lot, serial or model codes
- Serial Nos., UDI IPY0374, (01) 0 9342395 00012 0 ISRY0242, (01) 0 9342395 00015 1 ISG0370, (01) 0 9342395 00024 3 IPG0539, (01) 0 9342395 00009 0 BG1190
- 510(k) numbers
- ["K081565","K142398"]
- PMA numbers
- not on file
- Product code
- HQF
- Firm FEI
- 3002806902
- Firm city
- Lakes
- Firm state
- Mawson
- Firm country
- Australia
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28