FDA Medical Device Recalls (openFDA)
Mobius Mobility LLC: iBOT PMD with software version 01.05.24. Personal Mobility Device.
- Recall number
- Z-1965-2025
- Recalling firm
- Mobius Mobility LLC
- Product
- iBOT PMD with software version 01.05.24. Personal Mobility Device.
- Status
- Open, Classified
- Initiated
- 2025-04-24
- Posted
- 2025-06-12
- Terminated
- not on file
- Reason
- Software issue that could potentially lead to the device tipping over from Balance Mode.
- Root cause
- Software design
- Action
- Consignees were initially contacted via telephone on 4/24/25 to notify them of this software issue. An URGENT MEDICAL DEVICE CORRECTION NOTICE dated 5/5/25 was emailed to consignees to notify them of the recall and instructing them that the firm is developing a software correction. Once Mobius Mobility has obtained a correction they will reach out to consignees to schedule a time to update affected devices. Consignees with any questions can email service @mobiusmobility.com or call 833-346-4268x2.
- Quantity
- 442 units
- Distribution
- US Nationwide distribution in the states of CA, CO, FL, GA, IN, NC, NJ, OH, PR, RI, TN, TX & VA.
- Lot, serial or model codes
- Software Version 01.05.24; UDI (01) 00857584008010 (21) 0001108, (01) 00857584008010 (21) 0001110, (01) 00857584008010 (21) 0001121, (01) 00857584008010 (21) 0001123, (01) 00857584008010 (21) 0001128, (01) 00857584008010 (21) 0001136, (01) 00857584008010 (21) 0001137, (01) 00857584008010 (21) 0001422, (01) 00857584008010 (21) 0001438, (01) 00857584008010 (21) 0001496, (01) 00857584008010 (21) 0001505, (01) 00857584008010 (21) 0001508, (01) 00857584008010 (21) 0001516, (01) 00857584008010 (21) 0001517, (01) 00857584008010 (21) 0001519; Serial No. 0001108, 0001422, 0001438, 0001496, 0001505, 0001508, 0001516, 0001517, 0001519, 0001110, 0001121, 0001123, 0001128, 0001136, 0001137.
- 510(k) numbers
- ["K243442"]
- PMA numbers
- not on file
- Product code
- IMK
- Firm FEI
- 3014522447
- Firm city
- Manchester
- Firm state
- NH
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28