Toshiba American Medical Systems Inc: SSH-880CV Aplio Artida System

Recall number
Z-1955-2010
Recalling firm
Toshiba American Medical Systems Inc
Product
SSH-880CV Aplio Artida System
Status
Terminated
Initiated
2009-09-16
Posted
2010-07-06
Terminated
2012-03-21
Reason
Toshiba America Medical Systems (TAMS) is initiating a field correction on the SSH-880CV Aplio Artida System due to a problem related to a wrong measurement value display in using with DICOM viewer. When an image is transferred from this system in DICOM transfer and a measurement is performed on a Doppler image in the DICOM viewer, measured values such as VTI (Velocity Time Integral) and PI/RI
Root cause
Software design
Action
The software will be replaced. To accomplish this, Toshiba America Medical Systems (TAMS) will notify consignees via USPS return receipt mail. The letter contains a return reply form that is to be faxed to TAMS for retention. For questions or concerns, please contact your Toshiba representative or the firm at (714) 730-5000.
Quantity
22
Distribution
Nationwide ((CA, GA, IA, LA, MA, MD, MN, MT, PA, PR, OH, OR, TN).
Lot, serial or model codes
2031, 2032, 2033, 2048, 2049, 2056, 2069, 2070, 2071, 2072, 2073, 2074, 2075, 2076, 2077, 2078, 2079, 2080, 2087, 2088, 2103, 2104, 2105, 2106, 2707, 2108, 2119, 2120, 2123, 2124, 2204, 2205, 2206
510(k) numbers
["K090158"]
PMA numbers
not on file
Product code
IYO
Firm FEI
2020563
Firm city
Tustin
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register