FDA Medical Device Recalls (openFDA)
Toshiba American Medical Systems Inc: SSH-880CV Aplio Artida System
- Recall number
- Z-1955-2010
- Recalling firm
- Toshiba American Medical Systems Inc
- Product
- SSH-880CV Aplio Artida System
- Status
- Terminated
- Initiated
- 2009-09-16
- Posted
- 2010-07-06
- Terminated
- 2012-03-21
- Reason
- Toshiba America Medical Systems (TAMS) is initiating a field correction on the SSH-880CV Aplio Artida System due to a problem related to a wrong measurement value display in using with DICOM viewer. When an image is transferred from this system in DICOM transfer and a measurement is performed on a Doppler image in the DICOM viewer, measured values such as VTI (Velocity Time Integral) and PI/RI
- Root cause
- Software design
- Action
- The software will be replaced. To accomplish this, Toshiba America Medical Systems (TAMS) will notify consignees via USPS return receipt mail. The letter contains a return reply form that is to be faxed to TAMS for retention. For questions or concerns, please contact your Toshiba representative or the firm at (714) 730-5000.
- Quantity
- 22
- Distribution
- Nationwide ((CA, GA, IA, LA, MA, MD, MN, MT, PA, PR, OH, OR, TN).
- Lot, serial or model codes
- 2031, 2032, 2033, 2048, 2049, 2056, 2069, 2070, 2071, 2072, 2073, 2074, 2075, 2076, 2077, 2078, 2079, 2080, 2087, 2088, 2103, 2104, 2105, 2106, 2707, 2108, 2119, 2120, 2123, 2124, 2204, 2205, 2206
- 510(k) numbers
- ["K090158"]
- PMA numbers
- not on file
- Product code
- IYO
- Firm FEI
- 2020563
- Firm city
- Tustin
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28