Skip to the content

FDA Medical Device Recalls (openFDA)

Zimmer, Inc.: Natural Knee II Revision Femoral Stem, 125mm, item 6215-12-185. packaged in configuration 803/791. Knee Prosthesis component.

Recall number
Z-1948-2015
Recalling firm
Zimmer, Inc.
Product
Natural Knee II Revision Femoral Stem, 125mm, item 6215-12-185. packaged in configuration 803/791. Knee Prosthesis component.
Status
Terminated
Initiated
2014-10-02
Posted
2015-07-01
Terminated
2015-07-01
Reason
Failed to meet established requirements for sterile barrier integrity due to a breach of the inner cavity. If the outer tray is compromised in addition to the inner tray a periprosthetic infection may occur. The risk associated with a periprosthetic infection is a revision or potential multi-stage revision to treat the infection.
Root cause
Package design/selection
Action
On 10/2/2014, URGENT MEDICAL DEVICE RECALL notifications were sent to the affected distributors and hospital Risk Managers. All distributors were notified via electronic mail, and all hospital Risk Managers and distributors with affected inventory were also notified via courier. The recall notification included a description of the reason for the recall, affected product, and consignee responsibilities. If after reviewing this notification you have further questions or concerns please call the customer call center at 1-800-348-2759. Hours of operation are Monday through Friday, 8 a.m. through 8 p.m. ET.
Quantity
352
Distribution
US Nationwide Distribution, including DC and the states of AL, AZ, CA, CO, FL, GA, IL, IN, KS, LA, MD, MI, MN, MO, MT, NC, ND, NJ, NY, OH, OK, OR, PA, TN, TX, UT, VA, WA, WI, and WY.
Lot, serial or model codes
item 6215-12-185 Lots: 7884165 7884736 7884737 7884738 7890330 11011448 11012660 60394169 60415420 60515793 60564530 60601129 60614014 60797044 60803056 60841848 60922655 60922656 60927853 60989112 61017152 61104192 61135872 61200531 61284571 61366919 61464308 61505310 61533025 61566733 61742151 61780361 61784264 61834192 61834778 61840307 61930376 61947884 61947885 62097910 62146925 62195165 62249431 62281270 62305675 62317987 62358643 62448678 62536901 62536902 62536903 62546527 62560372 62587577 62610989 62624715 62624716 62624717 62624718 62709664 62709665 62709667 62709670
510(k) numbers
["K973412"]
PMA numbers
not on file
Product code
JWH
Firm FEI
1000220733
Firm city
Warsaw
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register