Skip to the content

FDA Medical Device Recalls (openFDA)

Zimmer, Inc.: Brand Name: Disposable Mixing Bowls with Spatula Product Name: Mixing Bowl and Spatula Model/Catalog Number: 00-5049-011-00 Software Version: N/A Product Description: Mixing Bowl and Spatula Component: N/A

Recall number
Z-1946-2026
Recalling firm
Zimmer, Inc.
Product
Brand Name: Disposable Mixing Bowls with Spatula Product Name: Mixing Bowl and Spatula Model/Catalog Number: 00-5049-011-00 Software Version: N/A Product Description: Mixing Bowl and Spatula Component: N/A
Status
Open, Classified
Initiated
2026-03-30
Posted
2026-04-24
Terminated
not on file
Reason
Ten complaints have been received identifying issues at the time of use related to the package seal, including incomplete seals, wrinkles in the seals or peeling seals. Compromised sterile barrier may lead to clinically insignificant extension of surgery or infection leading to surgical intervention.
Root cause
Vendor change control
Action
On March 30, 2026, the firm began notifying customers via letters titled "URGENT MEDICAL DEVICE RECALL." Risk Managers were instructed to notify affected personnel and locate and quarantine affected product in inventory. A Zimmer Biomet sales representative may remove and return the affected product on your behalf, or the customer may choose to directly return all affected product from their facility. Surgeons were informed that there are no specific patient-monitoring instructions recommended beyond existing follow up schedules.
Quantity
66960
Distribution
US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, FL, GA, IA, ID, IL, IN, KS, KY, LA, MD, MI, MO, MS, NC, NE, NM, NY, OH, OK, PA, SD, TN, TX, UT, WA, WI and the countries of Canada, Australia, New Zealand, Malaysia, EMEA, Chile.
Lot, serial or model codes
Lot Code: Model No. 00504901100 ; 20-Pack UDI (01)00889024379718(17)301014(10)85390029 ; Individual UDI (01)00889024376564(17)301014(10)85390029 ; Lot Number 85390029 Model No. 00504901100 ; 20-Pack UDI (01)00889024379718(17)301103(10)85682882 ; Individual UDI (01)00889024376564(17)301103(10)85682882 ; Lot Number 85682882 Model No. 00504901100 ; 20-Pack UDI (01)00889024379718(17)301001(10)85390028 ; Individual UDI (01)00889024376564(17)301001(10)85390028 ; Lot Number 85390028 Model No. 00504901100 ; 20-Pack UDI (01)00889024379718(17)300422(10)82395239 ; Individual UDI (01)00889024376564(17)300422(10)82395239 ; Lot Number 82395239 Model No. 00504901100 ; 20-Pack UDI (01)00889024379718(17)300423(10)82395240 ; Individual UDI (01)00889024376564(17)300423(10)82395240 ; Lot Number 82395240 Model No. 00504901100 ; 20-Pack UDI (01)00889024379718(17)300929(10)85390027 ; Individual UDI (01)00889024376564(17)300929(10)85390027 ; Lot Number 85390027 Model No. 00504901100 ; 20-Pack UDI (01)00889024379718(17)300924(10)85390026 ; Individual UDI (01)00889024376564(17)300924(10)85390026 ; Lot Number 85390026 Model No. 00504901100 ; 20-Pack UDI (01)00889024379718(17)300415(10)82395238 ; Individual UDI (01)00889024376564(17)300415(10)82395238 ; Lot Number 82395238 Model No. 00504901100 ; 20-Pack UDI (01)00889024379718(17)300922(10)85390025 ; Individual UDI (01)00889024376564(17)300922(10)85390025 ; Lot Number 85390025 Model No. 00504901100 ; 20-Pack UDI (01)00889024379718(17)300826(10)85390022 ; Individual UDI (01)00889024376564(17)300826(10)85390022 ; Lot Number 85390022 Model No. 00504901100 ; 20-Pack UDI (01)00889024379718(17)300918(10)85390024 ; Individual UDI (01)00889024376564(17)300918(10)85390024 ; Lot Number 85390024 Model No. 00504901100 ; 20-Pack UDI (01)00889024379718(17)300722(10)85390023 ; Individual UDI (01)00889024376564(17)300722(10)85390023 ; Lot Number 85390023 Model No. 00504901100 ; 20-Pack UDI (01)00889024379718(17)310122(10)86970970 ; Individual UDI (01)00889024376564(17)310122(10)86970970 ; Lot Number 86970970
510(k) numbers
not on file
PMA numbers
not on file
Product code
JDZ
Firm FEI
1000220733
Firm city
Warsaw
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28