Skip to the content

FDA Medical Device Recalls (openFDA)

Newport Corp: Newport s N-LHP-928(j) units if sold as a laser head/power supply set, and N5-LPL-944-080 and N6-LPL-944-080 if sold as a power supply alone.

Recall number
Z-1946-2024
Recalling firm
Newport Corp
Product
Newport s N-LHP-928(j) units if sold as a laser head/power supply set, and N5-LPL-944-080 and N6-LPL-944-080 if sold as a power supply alone.
Status
Open, Classified
Initiated
2024-03-27
Posted
2024-06-05
Terminated
not on file
Reason
According to 21 CFR 1040.10(f)(4), each laser system classified as a Class IIIb or IV laser product shall incorporate a key-actuated master control. The key shall be removable and the laser shall not be operable when the key is removed. However, the key of the subject laser products can be removed from key switch and the laser remains on. This is a safety concern.
Root cause
Radiation Control for Health and Safety Act
Action
Newport will contact each affected customer, distributor and end user, to determine the status of the laser s current hardware and whether the power supply must be replaced. If the power supply has the incorrect hardware and must be replaced, Newport will remedy the defect by replacing the unit with a compliant unit.
Quantity
19
Distribution
US Nationwide
Lot, serial or model codes
P/N N-LHP-928 Set S/N (laser head + power supply): 6905AE, 6907AE, 9023AE, 3148AF, 6899AE 6909AE, 9025AE, 5695AF, 5696AF, 5691AF, 5692AF, 5693AF, 5694AF (9 of them in Irvine warehouse) P/N: N6-LPL-944-080 Power supply S/N: 23060081, 23060079
510(k) numbers
not on file
PMA numbers
not on file
Product code
RFN
Firm FEI
3011118306
Firm city
Ivine
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register