Skip to the content

FDA Medical Device Recalls (openFDA)

Verathon, Inc.: The BladderScan BVI 9400 is a portable ultrasound instrument that provides a noninvasive measurement of urinary bladder volume. The device consists of an ultrasound probe that scans the patient's bladder, and a battery operated console that provides measurement related information. The BVI 9400 measures ultrasonic reflections on multiple planes inside the body and produces a three dimensional image. Based on the image, the BVI 9400 calculates and displays bladder volume.

Recall number
Z-1939-2009
Recalling firm
Verathon, Inc.
Product
The BladderScan BVI 9400 is a portable ultrasound instrument that provides a noninvasive measurement of urinary bladder volume. The device consists of an ultrasound probe that scans the patient's bladder, and a battery operated console that provides measurement related information. The BVI 9400 measures ultrasonic reflections on multiple planes inside the body and produces a three dimensional image. Based on the image, the BVI 9400 calculates and displays bladder volume.
Status
Terminated
Initiated
2009-05-18
Posted
2009-09-17
Terminated
2010-10-07
Reason
Three specific issues are being addressed: 1. Calculation and use of an incorrect year in dates after 12/31/2009. 2. Double scans, continuous scans, and double printing. 3. Calibration sensitivity.
Root cause
Software design
Action
U.S. customers (including Government) were sent a post card on or about 5/18/2009 and were called about the upgrade. Units are to be upgraded by the customers using an online tool called ScanPoint. Customers may elect to return their units to the firm for upgrading. Canadian customers were phoned by Verathon. For foreign units sold and distributed through one of the recalling firm's related firms, Service Bulletin SB-0014 was sent to the distributors located in The Netherlands, United Kingdom, and Australia. Contact Verathon Customer Care at 1-800-331-2313 if you have any questions.
Quantity
1,978
Distribution
Worldwide - US, Canada, Malaysia, Hong Kong, Singapore, Argentina, New Zealand, Barbados, Dominican Republic, and Costa Rica.
Lot, serial or model codes
Units with serial numbers ranging from 1000 to 3179 are affected. (Serial number 1564 was not used.) The serial number is assigned as a sequential number, with no breakdown.
510(k) numbers
K071217
PMA numbers
not on file
Product code
IYO
Firm FEI
3022472
Firm city
Bothell
Firm state
WA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register