Abbott Molecular: VP 2000 Processor; an automated tissue stainer 100 volt - part 30-144100 or list 2J11-60; 117 volt - part 30-144101 or list 2J11-01, and part 30-102832 - processor only; 230 volt - part 30-144102 or list 2J11-04, and part 30-102833 processor only. Designed to automate and standardize slide specimen processing including deparafinization, FISH pretreatment and routine slide staining for the laboratory. The unit is preprogrammed to perform slide pretreatment protocols for FISH assays and can be programmed to perform routine histology/cytology slide staining or routine slide washing according to user requirements.

Recall number
Z-1929-2009
Recalling firm
Abbott Molecular
Product
VP 2000 Processor; an automated tissue stainer 100 volt - part 30-144100 or list 2J11-60; 117 volt - part 30-144101 or list 2J11-01, and part 30-102832 - processor only; 230 volt - part 30-144102 or list 2J11-04, and part 30-102833 processor only. Designed to automate and standardize slide specimen processing including deparafinization, FISH pretreatment and routine slide staining for the laboratory. The unit is preprogrammed to perform slide pretreatment protocols for FISH assays and can be programmed to perform routine histology/cytology slide staining or routine slide washing according to user requirements.
Status
Terminated
Initiated
2008-09-15
Posted
2009-08-30
Terminated
2010-01-07
Reason
Thermal runaway conditions, in that the target temperature was exceeded and the temperature display could not track the actual temperature within the heated slide basins of the VP 2000 Slide Processor.
Root cause
Device Design
Action
Abbott sent Device Recall/Urgent Field Safety Notice letters dated 9/15/08 to the VP2000 customers on 9/12/08, informing them that thermal runaway conditions may occur within the heated slide basins of the VP2000 Slide Processor. The customers were given actions to take to prevent this condition: a) VP2000 should be powered down when not in use and when the controlling computer is not in use; b) Liquid level in all heated basins should be checked periodically to ensure that there is adequate reagent present for slide processing and to keep the sensor probe immersed for accurate temperature sensing; and c) If large, rapid and random fluctuations in indicated basin temperature occurs the unit should be shut down and a Field Service Representative be contacted. Any questions were directed to their local Abbott Technical Service Representative.
Quantity
261 units
Distribution
Worldwide Distribution -- USA, Australia, Austria, Belgium, Canada, Columbia, Czech Republic, Egypt, France, Germany, Hong Kong, India, Ireland, Italy, Japan, Korea, Malaysia, Norway, Russia, Saudi Arabia, Singapore, Switzerland and the United Kingdom.
Lot, serial or model codes
all units
510(k) numbers
not on file
PMA numbers
not on file
Product code
KEY
Firm FEI
3005248192
Firm city
Des Plaines
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
3e6df46828c1
Size
288213574 bytes
Published
2026-10-06
Read
2026-10-07T03:20:25.099Z
Source file
Source file retained · verify
Refresh due
2026-10-14

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 3e6df46828c1, published 2026-10-06

Search this source