American Contract Systems, Inc.: LOWER EXTREMITY, Model: NCEX20I; LOWER EXTREMITY, Model: NCEX20J; TOTAL KNEE PACK, Model: NCTK31K; TOTAL KNEE PACK, Model: NCTK31L All convenience kits contain Specialist Blend Cast Padding, Item: QTX30-322NS
- Recall number
- Z-1916-2025
- Recalling firm
- American Contract Systems, Inc.
- Product
- LOWER EXTREMITY, Model: NCEX20I; LOWER EXTREMITY, Model: NCEX20J; TOTAL KNEE PACK, Model: NCTK31K; TOTAL KNEE PACK, Model: NCTK31L All convenience kits contain Specialist Blend Cast Padding, Item: QTX30-322NS
- Status
- Open, Classified
- Initiated
- 2025-04-21
- Posted
- 2025-06-06
- Terminated
- not on file
- Reason
- Test results for cast padding component included in procedure trays shows Ethylene Oxide/Ethylene Chlorohydrin (EO/ECH) residuals exceeded residual limits, which may lead to a possible loss or lack of functionality that may lead to a delay or prolonged treatment/procedure/therapy.
- Root cause
- Finished device change control
- Action
- On 4/21/2025, correction notices were emailed to the customer who was asked to do the following: 1) Discontinue use of the Cast Padding component. All other components within the kit can be used. 2) Add warning labels to affected kits in Table 2 to ensure the component is discarded. A labeling template is enclosed for printing labels which can be affixed on all affected kits in inventory. This label should be applied to each affected Procedure tray and product case in a prominently visible location to end-users. The location chosen should not cover any other critical product information found on existing product labeling. - Customers (end-users): If you received product from a distributor, please contact the distributor or your sales representative with any questions you may have. - Distributors: If any of the affected procedure trays identified above were further distributed, please ensure all end-users are appropriately notified of this product correction and maintain records of effectiveness. You are also responsible for the reconciliation of these customers. Please do not have your customers respond directly to ACS. Complete and return the response form via email to GM-OMRA_RECALLS@owens-minor.com Report any quality problems or adverse events associated with the products listed in this notification to complaints@owens-minor.com
- Quantity
- 850 Kits
- Distribution
- US Nationwide distribution in the state of CO.
- Lot, serial or model codes
- Model/UDI-DI/Lot(Expiration): NCEX20I/00191072222398/9146304(2025-12-03), 9207685(2026-01-08); NCEX20J/00191072234582/9222096(2026-01-30), 9222096(2026-01-30); NCTK31K/00191072222480/9086225(2025-11-01), 9086225(2025-11-01); NCTK31L/00191072234872/9219596(2026-01-15), 9219596(2026-01-15)
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- OJH
- Firm FEI
- 3010126341
- Firm city
- Chandler
- Firm state
- AZ
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28