Skip to the content

FDA Medical Device Recalls (openFDA)

Synthes (USA) Products LLC: Synthes Torque Limiting Handle, Part Number 03.231.013 This device is used in systems indicated for fracture fixation.

Recall number
Z-1916-2018
Recalling firm
Synthes (USA) Products LLC
Product
Synthes Torque Limiting Handle, Part Number 03.231.013 This device is used in systems indicated for fracture fixation.
Status
Terminated
Initiated
2018-05-04
Posted
not on file
Terminated
2019-10-21
Reason
The adapter may demonstrate a degradation of function as the final products were assembled with ball bearings of high carbon content that is likely to rust prematurely.
Root cause
Component design/selection
Action
An Urgent Medical Device Recall Notification letters dated May 4, 2018, were distributed to customers. The letter requests that customers take the following actions: Potential Patient Impact: The subject, Torque Limiting Handle, may demonstrate a degradation of function with the adapter collar sticking, and/or not retaining the mating adapter male shaft of the driver component. In addition, the ball bearings may corrode over time causing a failure in the torque limiting feature. The following potential scenarios could occur: " Malunion/nonunion and device loosening if there is a degradation of function and too much or too little torque is applied. " Adverse tissue reaction if there is corrosion and any debris enters an open incision. " Surgical delay due to the time required to investigate and decide next steps if the instrument fails during surgery. If you have no product to be returned: " Return the Verification Section (page 3 of this letter) to DePuy Synthes by: o Fax: 866-853-1969 or o Scan/email: Synthes7370@stericycle.com NOTE: If the Verification Section is completed on behalf of more than one facility and/or individual, please clearly indicate the name and address of the facility and/or individual in page number three (3) of this notification. 3. Forward this notice to anyone in your facility that needs to be informed. 4. If any of the subject product has been forwarded to another facility, contact that facility to arrange return. 5. Keep a copy of this notice visibly posted for awareness until all product subject to this recall (removal) have been returned. While processing your returns, please maintain a copy of this notice with the product subject to this recall (removal) and keep a copy for your records. For further questions, please call (610) 719-5450.
Quantity
50
Distribution
USA (nationwide) Distribution to the states of : AK, CA, FL, ID, IN, KS, KY, LA, MD, MI, MT, NC, NJ, NY, OH, OK, PA, SD, TN, TX, UT, WI, and WV.
Lot, serial or model codes
H344006-02,  H344006-03,  H344006-04,  H344006-05,  H344006-06,  H344006-07,  H344006-08,  H344006-09,  H344006-10,  H344006-11,  H344006-12,  H344006-13,  H344006-14,  H344006-15,  H344006-16,  H344006-17,  H344006-18,  H344006-19,  H344006-20,  H344006-21,  H344006-22,  H344006-23,  H344006-24,  H344006-25,  H344006-26,  H344006-27,  H344006-28,  H344006-29,  H344006-30,  H344006-31,  H344006-32,  H344006-33,  H344006-34,  H344006-35,  H344006-36,  H344006-37,  H344006-38,  H344006-39,  H344006-40,  H344006-41,  H344006-43,  H344006-45,  H344006-46,  H344006-47,  H344006-48,  H344006-49,  H369323-03,  H369323-04
510(k) numbers
not on file
PMA numbers
not on file
Product code
HXX
Firm FEI
3008812560
Firm city
West Chester
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28