Skip to the content

FDA Medical Device Recalls (openFDA)

Olympus Corporation of the Americas: InstaClear Lens Sheath. The InstaClear lens cleaner system and versatile sheath cleans and clears the lens of a rigid endoscope almost instantly during surgery without have to remove it from the nasal cavity.

Recall number
Z-1888-2024
Recalling firm
Olympus Corporation of the Americas
Product
InstaClear Lens Sheath. The InstaClear lens cleaner system and versatile sheath cleans and clears the lens of a rigid endoscope almost instantly during surgery without have to remove it from the nasal cavity.
Status
Open, Classified
Initiated
2024-04-03
Posted
2024-05-23
Terminated
not on file
Reason
Potential for distal end of the sheath to break off during a procedure.
Root cause
Under Investigation by firm
Action
Consignees were mailed an URGENT MEDICAL DEVICE CORRECTION notification, dated 5/1/24. Consignees are instructed to read the provided labeling addendum, inspect their inventory to identify affected devices and place a copy of the addendum with the remaining inventory, ensure all required personnel are aware of this recall, notify those who received product further distributed, and acknowledge receipt of the recall notification using Olympus's recall portal. Consignees can report the incidence of a device breakage associated with this event to the Technical Assistance Center at 1-800-8489024, Option 1. Consignees with any questions can contact Olympus at 647-999-3203 or by email at Cynthia.Ow@olympus.com.
Quantity
pending
Distribution
Domestic: Nationwide Distribution; Foreign: Australia, Canada, Germany, Japan.
Lot, serial or model codes
Model No. LCS4K30BTOL, LCS1530TPOL, LCS1830TPST, LCS4K30TPOL, LCS1545BTOL, LCS1845BTST, LCS4K45BTOL, LCS1545TPOL, LCS1845TPST, LCS4K45TPOL, LCS1570BTOL, LCS1870BTST, LCS4K70BTOL, LCS1570TPOL, LCS1870TPST, LCS4K70TPOL; UDI-DI: 00821925040137, 00821925040151, 00821925040144, 00821925040175, 00821925040168, 00821925040199, 00821925040182, 00821925040083, 00821925040076, 00821925040106, 00821925040090, 00821925040120, 00821925040113, 00821925040069; All Lot Numbers.
510(k) numbers
["K152531"]
PMA numbers
not on file
Product code
EOB
Firm FEI
2429304
Firm city
Center Valley
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register