Cordis Corporation: Cordis OPTEASE Retrieval Inferior Vena Cava (IVC) Filter***REF Cat. No.***466F220A and 466F220B***Vena Cava Filter and Introduction Kit***(BRITE TIP Catheter Sheath Introducer and Angiographic Vessel Dilator)***Cordis***a Johnson and Johnson Company***ENDOVASCULAR*** Label No. SB 10247519.2/***LB466F220AO.5***omoraoro/2012.050794***Do not resterilize***. Product Usage: The OPTEASE Vena Cava Filter is indicated for use in the prevention of recurrent pulmonary embolism (PE) via percutaneous placement in the vena cava in several situations: Pulmonary thromboembolism when anticoagulants are contraindicated; Failure of anticoagulant therapy for thromboembolic disease; Emergency treatment following massive pulmonary embolism where anticipated benefits of conventional therapy are reduced and Chronic, recurrent pulmonary embolism where anticoagulant therapy has failed, or is contraindicated.
- Recall number
- Z-1887-2013
- Recalling firm
- Cordis Corporation
- Product
- Cordis OPTEASE Retrieval Inferior Vena Cava (IVC) Filter***REF Cat. No.***466F220A and 466F220B***Vena Cava Filter and Introduction Kit***(BRITE TIP Catheter Sheath Introducer and Angiographic Vessel Dilator)***Cordis***a Johnson and Johnson Company***ENDOVASCULAR*** Label No. SB 10247519.2/***LB466F220AO.5***omoraoro/2012.050794***Do not resterilize***. Product Usage: The OPTEASE Vena Cava Filter is indicated for use in the prevention of recurrent pulmonary embolism (PE) via percutaneous placement in the vena cava in several situations: Pulmonary thromboembolism when anticoagulants are contraindicated; Failure of anticoagulant therapy for thromboembolic disease; Emergency treatment following massive pulmonary embolism where anticipated benefits of conventional therapy are reduced and Chronic, recurrent pulmonary embolism where anticoagulant therapy has failed, or is contraindicated.
- Status
- Terminated
- Initiated
- 2013-03-29
- Posted
- 2013-08-07
- Terminated
- 2014-10-27
- Reason
- Correction to provide clarification and additional labeling to minimize likelihood on implanting filter backwards.
- Root cause
- Labeling design
- Action
- Cordis Corporation sent an "URGENT Medical Device Correction" notification to their consignees in the US by letter dated April 3, 2013 and an e-mail was sent to the Canadian consignees. The letter identified the affected product, problem and actions to be taken. Consignees were instructed to read the description and recommendations sections carefully, then sign and return the enclosed Acknowledgement Form directly to Cordis to their fax number(908) 429-8287 or Scan and e-mail to MiamiQA@crdus.jnj.com. The letter states to pass on the notice to anyone in the consignees' facilities that need to be informed. Additionally, the notification request consignees to maintain awareness of this notification until the information has been incorporated in the Cordis OPTEASE Vena Cava Filter labeling. For questions related to the issue contact Cordis Endovascular Medical Information Line 877-338-4235.
- Quantity
- 33,000 units
- Distribution
- Worldwide Distribution - USA Nationwide in the states of AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, Ny, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, including Hawaii and Puerto Rico and the countries of Canada & Panama
- Lot, serial or model codes
- Model Numbers 466F220A and 466F220B.
- 510(k) numbers
- ["K034050","K091077"]
- PMA numbers
- not on file
- Product code
- DTK
- Firm FEI
- 1016427
- Firm city
- Miami Lakes
- Firm state
- FL
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28