Skip to the content

FDA Medical Device Recalls (openFDA)

Stryker Endoscopy: Shaver Handpiece Sterilization Tray. Model number 272-700-000. For sterilization of Stryker Endoscopy arthroscopic shaver.

Recall number
Z-1872-2014
Recalling firm
Stryker Endoscopy
Product
Shaver Handpiece Sterilization Tray. Model number 272-700-000. For sterilization of Stryker Endoscopy arthroscopic shaver.
Status
Terminated
Initiated
2014-05-19
Posted
2014-06-25
Terminated
2015-05-07
Reason
STERRAD 100S parameters provided in the Shaver Handpiece reprocessing guide (P10299 Revision A) did not fully consider all worst-case scenarios during validation testing.
Root cause
Nonconforming Material/Component
Action
Device Correction letters were sent on May 23, 2014 to all users by certified mail. Internation sites will receive notification and be responsible for contacting customers.
Quantity
1992 trays
Distribution
Worldwide Distribution.
Lot, serial or model codes
Part Number 272-700-000: Lot number: 20476.1 23541 23837 26403 26404 26405 26676 27009 27518 28341 30248 34355 35932 36756 39705 05J034834 06con05774 06CON06360 06CON06846 06con06869 06con06960 06con08587 25067-1-12 26882-1-1 26882-1-11 26882-1-12 26882-1-13 26882-1-14 35200-1-1 36610-1-1 38820-1-1 39078-1-1 40504-2-1 40675-1-1 41231-1-1 41827-1-1 42085-1-1 42375-1-1 42527-1-1 42789-1-1 43103-1-1 43589-1-1 43716-1-1 43716-1-3 43716-1-4 44568-1-10 44568-1-11 44568-1-2 44568-1-3 44568-1-4 44568-1-6 44568-1-7 44568-1-8 44568-1-9 46227-1-1 47731-1-1 48595-1-1 48596-1-1 48597-1-1 48598-1-1 48599-1-1 48600-1-1 50101-1-1 50628-1-1 50828-1-1 51237-1-1 52439-1-1 55035-1-1 55422-1-1 55462-1-1 56456-1-1 R471611 SS15329.
510(k) numbers
["K941333"]
PMA numbers
not on file
Product code
HRX
Firm FEI
2936485
Firm city
San Jose
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28