Skip to the content

FDA Medical Device Recalls (openFDA)

Straumann USA LLC: Plus Initial Drill, Article Number: 103.170;

Recall number
Z-1861-2025
Recalling firm
Straumann USA LLC
Product
Plus Initial Drill, Article Number: 103.170;
Status
Open, Classified
Initiated
2025-04-14
Posted
2025-05-28
Terminated
not on file
Reason
The referred batch was produced without the laser engraving of 7 mm.
Root cause
Under Investigation by firm
Action
On April 14, 2025 customers were notified via phone with a follow-up letter sent on April 28, 2024. Both contacts included the following: Action to be taken: 1. If article 103.170 from lot LNWT8 is still in your inventory, then stop use/distribution of the product immediately and quarantine/segregate physically. 2. If the product was used and no complications or problems were identified, of the patient did not present symptoms there is no need for any additional action with the patient. 3. If article 103.170 from Lot LNWT8 is still in your inventory, return it for a credit or replacement as indicated on the Customer Confirmation Form. 4. In the product is not found in your stock, indicate this on the enclosed Customer Confirmation Form. 5 In all cases, complete and return the enclosed Customer Confirmation Form with the affected product using the return UPS label provided to: Straumann USA, LLC; 60 Minuteman Road, Andover, MA 01810. transmission of the Field Safety Notice This notice needs to be passed on to all those who need to be aware within your organization or to any organization where the affected devices have been transferred. The FDA has been informed about this Field Safety Corrective Action. We apologize for any inconvenience that this may cause.
Quantity
288 units ($& US, 241 OUS)
Distribution
Worldwide - US Nationwide distribution in the states of AZ, CA, FL, GA, IL, IN, MN, NC, NV, NY, OH, OK, SC, TX, VA, WI and the countries of Brazil, Germany, India, Indonesia, Portugal, Romania, Spain, Switzerland, Ukraine.
Lot, serial or model codes
Article Number: 103.170; UDI-DI: (01)7898237561967(11)250204(17)400204(10)LNWT8; Lot Number: LNWT8
510(k) numbers
not on file
PMA numbers
not on file
Product code
DZA
Firm FEI
1000121052
Firm city
Andover
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register