Varian Medical Systems Inc: The VariSource iX series afterloader systems are computer controlled remote electro/mechanical systems used for medical purposes, for placing a cable incorporating an irradiated iridium seed internally or close by a malignant tumor or tumor bed in a practice known as brachytherapy.

Recall number
Z-1849-2015
Recalling firm
Varian Medical Systems Inc
Product
The VariSource iX series afterloader systems are computer controlled remote electro/mechanical systems used for medical purposes, for placing a cable incorporating an irradiated iridium seed internally or close by a malignant tumor or tumor bed in a practice known as brachytherapy.
Status
Terminated
Initiated
2015-05-18
Posted
2015-06-26
Terminated
2016-09-12
Reason
After an application freeze and restart of the VariSource iX series by power cycling, the Partial Fraction generated by the system will not be correct. The application freeze issue affects only the systems equipped with magnetic Hard Disk Drives (HDD).
Root cause
Software Design Change
Action
All the consignees affected by this recall will be notified by: The Urgent Medical Device Correction Letter/ Field Safety Notification will be distributed to all affected users, with a description of the problem. The Letter will also be distributed to the Varian Sales, Marketing, and Service organizations, informing them of the issue. Additionally, the letter will be posted to the customer support site: http:/ /www.MyVarian.com.
Quantity
55
Distribution
Worldwide Distribution.
Lot, serial or model codes
JAQ
510(k) numbers
K122236
PMA numbers
not on file
Product code
JAQ
Firm FEI
3011400544
Firm city
Charlottesville
Firm state
VA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register