Skip to the content

FDA Medical Device Recalls (openFDA)

SteriGear, LLC: SteriGear Urinary Drain Bag with Fig Leaf Cover, 2000 ml, REF 10270, The Fig Leaf, SteriGear 362 S. University Avenue, provo, UT 84601, 1-800-398-3259, SteriGear.com, Sterile, EO Urinary Drainage Collection Kit, for Indwelling Catheter. This device is intended to collect the Urine drained from an indwelling catheter

Recall number
Z-1848-2014
Recalling firm
SteriGear, LLC
Product
SteriGear Urinary Drain Bag with Fig Leaf Cover, 2000 ml, REF 10270, The Fig Leaf, SteriGear 362 S. University Avenue, provo, UT 84601, 1-800-398-3259, SteriGear.com, Sterile, EO Urinary Drainage Collection Kit, for Indwelling Catheter. This device is intended to collect the Urine drained from an indwelling catheter
Status
Terminated
Initiated
2011-01-11
Posted
2014-06-23
Terminated
2015-01-02
Reason
SteriGear notifed customers of circumstances in which The Fig Leaf urinary drain bag valve can close creating a negative pressure that can suck the non-return valve closed.
Root cause
Device Design
Action
SteriGear, LLC sent a special note to all affected customers between January 12, 2011 and September 28, 2012. The note identified the product, the problem, and the action to be taken by the customer. Customers were instructed on how to reposition the inlet tube. Customers with questions were instructed to contact Director of Product Development at 801-607-3211. No correction or removal was performed. For questions regarding this recall call 801-607-3209.
Quantity
110,560
Distribution
Nationwide Distribution
Lot, serial or model codes
UPD 851616002007, Part 10270
510(k) numbers
not on file
PMA numbers
not on file
Product code
FCN
Firm FEI
3007608536
Firm city
Provo
Firm state
UT
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register