Skip to the content

FDA Medical Device Recalls (openFDA)

Biomet, Inc.: Various trauma and sports medicine instruments and implants. Orthopedic manual surgical instrument.

Recall number
Z-1827-2016
Recalling firm
Biomet, Inc.
Product
Various trauma and sports medicine instruments and implants. Orthopedic manual surgical instrument.
Status
Terminated
Initiated
2016-04-01
Posted
2016-05-26
Terminated
2017-05-23
Reason
Insufficient seal strength on sterile barrier packaging. The seal could be jeopardized causing the sterile barrier to be compromised and the product could become nonsterile. If the product becomes non-sterile and is used or implanted, it may lead to an infection.
Root cause
Equipment maintenance
Action
Zimmer Biomet sent an Urgent Medical Device Recall Notice dated April 1, 2016, to all affected customers. Customers were instructed to locate and return the affected product, and implanting surgeons were instructed to monitor patients. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. Questions related to this notice should be directed to (574) 372-1570, Monday through Friday, 8 a.m. to 5 p.m.
Quantity
not on file
Distribution
Worldwide Distribution - US (Nationwide) and globally to Argentina, Australia, Brazil, Canada, Chile, Colombia, Costa Rica, Ecaudor, The Netherlands, Japan, Korea, Malaysia, Mexico, Peru.
Lot, serial or model codes
Part #'s 900728 900737 900738 900740 909836 110008343 110009769 231201303  Lot #'s 669220 625160 625180 679060 529190 585920 742460 751280 751290 751330 751400 751410 751420 751430 630570
510(k) numbers
not on file
PMA numbers
not on file
Product code
LXH
Firm FEI
1825034
Firm city
Warsaw
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register