Skip to the content

FDA Medical Device Recalls (openFDA)

Onkos Surgical, Inc.: ELEOS COLLAR STEM, CEMENTED, FLUTED, 13MM X 120MM. Model Number: HC-13120-03M. Component of Limb Salvage System with BioGrip.

Recall number
Z-1817-2025
Recalling firm
Onkos Surgical, Inc.
Product
ELEOS COLLAR STEM, CEMENTED, FLUTED, 13MM X 120MM. Model Number: HC-13120-03M. Component of Limb Salvage System with BioGrip.
Status
Open, Classified
Initiated
2023-12-22
Posted
2025-05-23
Terminated
not on file
Reason
Mislabeling of 13mm Modular Segmental Stem package (HC-13120-03M) which included an implant that was 15mm in diameter (HC-15120-03M) instead of being 13mm as stated on the package. The corresponding 15mm cemented stems were mislabeled as size 13mm cemented stems.
Root cause
Labeling Change Control
Action
ONKOS Surgical notified their distributors on about 12/22/2023 via email. They were instructed to segregate and quarantine the product immediately, complete and return the "Product and Distribution Information" form within 2 business days and return any existing implants on hand within 2 business days after signing the acknowledgement. They were also instructed that they will receive replacement implants and to place them in their modular segmental collar stem implant packs.
Quantity
23 units
Distribution
US Nationwide distribution in the states of California, Kentucky, and Wisconsin.
Lot, serial or model codes
Model Number: HC-13120-03M. UDI-DI: B278HC1312003M0. Serial Numbers: P230274-1XX (XX represents 23 parts in sequential order from 101-123).
510(k) numbers
["K203815"]
PMA numbers
not on file
Product code
JDI
Firm FEI
3013450937
Firm city
Parsippany
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register