Skip to the content

FDA Medical Device Recalls (openFDA)

Intuitive Surgical, Inc.: INTUITIVE SURGICAL DA VINCI SURGICAL SYSTEM, MODE; S Cardiac Probe Gasper Instrument, Model IS2000, Part Number 420215-02; Product is manufactured and distributed by Intuitive Surgical, Inc. Sunnyvale, CA . The IS2000 CPG is used during cardiac ablation procedures.

Recall number
Z-1811-2008
Recalling firm
Intuitive Surgical, Inc.
Product
INTUITIVE SURGICAL DA VINCI SURGICAL SYSTEM, MODE; S Cardiac Probe Gasper Instrument, Model IS2000, Part Number 420215-02; Product is manufactured and distributed by Intuitive Surgical, Inc. Sunnyvale, CA . The IS2000 CPG is used during cardiac ablation procedures.
Status
Terminated
Initiated
2008-04-18
Posted
2008-09-17
Terminated
2011-03-25
Reason
The product has a software interface problem. When the product is connected to the da Vinci IS2000 System it will not allow the System to recognize the Instrument which makes the IS2000 CPG instrument nonfunctional at all sites. Risks associated are loss of operability of the instrument, delay in surgery, and loss of dexterity.
Root cause
Software design
Action
On April 18, 2008 an Urgent Device Recall letter was issued to customers. Consignees were notified via e-mail and were instructed to return the product. Customers were instructed to segregate the product in a secure area for Intuitive Surgical CSRs. If you have any questions contact Karen Uyesugi at (408) 523-2100.
Quantity
11
Distribution
Class II Recall - Nationwide Distribution --- including states of NC, OH, GA and WI.
Lot, serial or model codes
Lot Number: 802011
510(k) numbers
["K070947"]
PMA numbers
not on file
Product code
NAY
Firm FEI
3001675293
Firm city
Sunnyvale
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register