FDA Medical Device Recalls (openFDA)
MEDLINE INDUSTRIES, LP - Northfield: Medline Kits containing Codman Disposable Perforator 14 mm: 1. CRANI ADD A PACK, SKU DYNJ30926O 2. CRANIOTOMY, SKU CDS983611N 3. CRANIOTOMY CDS, SKU CDS983467J 4. CRANIOTOMY PACK, SKU DYNJ85807A, DYNJ85927 5. MAJOR CRANIOTOMY PACK, SKU DYNJ82007A
- Recall number
- Z-1809-2025
- Recalling firm
- MEDLINE INDUSTRIES, LP - Northfield
- Product
- Medline Kits containing Codman Disposable Perforator 14 mm: 1. CRANI ADD A PACK, SKU DYNJ30926O 2. CRANIOTOMY, SKU CDS983611N 3. CRANIOTOMY CDS, SKU CDS983467J 4. CRANIOTOMY PACK, SKU DYNJ85807A, DYNJ85927 5. MAJOR CRANIOTOMY PACK, SKU DYNJ82007A
- Status
- Open, Classified
- Initiated
- 2025-04-22
- Posted
- 2025-05-31
- Terminated
- not on file
- Reason
- Affected kits contain recalled Integra Lifesciences Codman Disposable Perforators due to inadequate welds on specific lots. Perforators with inadequate welds may break or separate.
- Root cause
- Nonconforming Material/Component
- Action
- On April 22, 2025, the firm notified customers of the recall through letters titled Immediate Action Required Voluntary Medical Device Recall. Customers were instructed to identify and quarantine all affected product on hand. Customers are to submit a response form to the recalling firm. Upon receipt of the response firm, Medline will provide the customer with over-labels to place on affected inventory, which will include instructions for staff to remove and discard the affected component prior to using the kit. Distributors should notify downstream customers/accounts of this recall. If you have any questions, contact the Recall Department at 866-359-1704 or recalls@medline.com.
- Quantity
- 1853 total
- Distribution
- US distribution to CA, FL, IL, IN, KY, MA, MD, MN, NY, TX, VT, WA and WI. No OUS distribution.
- Lot, serial or model codes
- 1. CRANI ADD A PACK, SKU DYNJ30926O, UDI-DI 10195327539153 (EA), 40195327539154 (CASE) KIT LOTS 24AMF811 2. CRANIOTOMY, SKU CDS983611N, UDI-DI 10195327466626 (EA), 40195327466627 (CASE) KIT LOTS 23KBA720 23LBI611 23LBJ586 3. CRANIOTOMY CDS, SKU CDS983467J, UDI-DI 10195327561116 (EA), 40195327561117 (CASE) KIT LOTS 24ADA818 24ADC216 24BDA929 24DDB447 24EDA275 24EDA751 24GDB213 24IDA327 24IDB043 4. CRANIOTOMY PACK, SKU DYNJ85807A, UDI-DI 10195327637392 (EA), 40195327637393 (CASE) KIT LOTS 24CBB843 24EBC889 24GBY029 ; SKU DYNJ85927, UDI-DI 10195327575571 (EA), 40195327575572 (CASE) KIT LOTS 24BMD310 24EMI836 5. MAJOR CRANIOTOMY PACK, SKU DYNJ82007A UDI-DI 10195327635466 (EA), 40195327635467 (CASE) KIT LOTS 24CMA496 24HMI081 24HMI978
- 510(k) numbers
- K213481
- PMA numbers
- not on file
- Product code
- LRO
- Firm FEI
- 1417592
- Firm city
- Northfield
- Firm state
- IL
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28