Skip to the content

FDA Medical Device Recalls (openFDA)

Beckman Coulter, Inc.: MicroScan Neg Multidrug Resistant MIC 1 REF C32368 UDI-DI code: 15099590720568 Intended Use- For use with MicroScan Dried Gram Negative MIC/Combo Panels and Dried Gram Negative Breakpoint Combo Panels. MicroScan panels are designed for use in determining antimicrobial agent susceptibility and/or identification to the species level of aerobic and facultatively anaerobic gram-negative bacilli.

Recall number
Z-1794-2026
Recalling firm
Beckman Coulter, Inc.
Product
MicroScan Neg Multidrug Resistant MIC 1 REF C32368 UDI-DI code: 15099590720568 Intended Use- For use with MicroScan Dried Gram Negative MIC/Combo Panels and Dried Gram Negative Breakpoint Combo Panels. MicroScan panels are designed for use in determining antimicrobial agent susceptibility and/or identification to the species level of aerobic and facultatively anaerobic gram-negative bacilli.
Status
Open, Classified
Initiated
2025-09-08
Posted
2026-04-08
Terminated
not on file
Reason
Due to manufacturing issue (incorrectly processing) with in vitro multidrug resistant test.
Root cause
Under Investigation by firm
Action
On 09/08/2025, the firm sent via mail/email an "URGENT FIELD SAFETY NOTICE" to customers informing them that as a result of an internal investigation Beckman Coulter confirmed that Ciprofloxacin (Cp) antimicrobial agent was incorrectly manufactured on MicroScan Neg Multidrug Resistant MIC 1 Lot 2026-05-08. Customers are instructed to: "Review Ciprofloxacin results generated from Lot 2026-05-08 to identify any abnormal or unexpected trends in both susceptible and resistance patterns, or unexpected change in MIC trends, as compared to historical results for that same species (e.g. a sudden, unexplained increase in susceptible or % susceptible results for a species after implementation of Lot 2026-05-08 versus the previous susceptible or % susceptible results for the same species prior to using Lot 2026-05-08). "Discontinue use of this lot and discard per your laboratory guidelines "The laboratory should retain the inventory of other MicroScan Neg Multidrug Resistant MIC 1 lots as they are not impacted by this issue. Note-Beckman Coulter is no longer distributing MicroScan Neg Multidrug Resistant MIC 1 Lot 2026-05-08 For questions - contact Customer Support Center from website: http://www.beckmancoulter.com
Quantity
3 boxes
Distribution
International distribution to the country of Poland.
Lot, serial or model codes
Catalog Number: C32368 UDI-DI code: 15099590720568 Lot Number: 2026-05-08
510(k) numbers
not on file
PMA numbers
not on file
Product code
LTT
Firm FEI
2919016
Firm city
West Sacramento
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register