Myoscience Inc: Myoscience brand iovera Smart Tip, Myoscience Cryo-Touch IV, a Cryogenic Surgical device, REF Catalog Number: STA0311-10. Product Usage: The Myoscience Cryo-Touch IV (a.k.a. iovera) is used to destroy tissue during surgical procedures by applying freezing cold. It can also be used to produce lesions in peripheral nervous tissue by the application of cold to the selected site for the blocking of pain. The Cryo-Touch IV (a.k.a. iovera) is not indicated for treatment of central nervous system tissue.

Recall number
Z-1781-2013
Recalling firm
Myoscience Inc
Product
Myoscience brand iovera Smart Tip, Myoscience Cryo-Touch IV, a Cryogenic Surgical device, REF Catalog Number: STA0311-10. Product Usage: The Myoscience Cryo-Touch IV (a.k.a. iovera) is used to destroy tissue during surgical procedures by applying freezing cold. It can also be used to produce lesions in peripheral nervous tissue by the application of cold to the selected site for the blocking of pain. The Cryo-Touch IV (a.k.a. iovera) is not indicated for treatment of central nervous system tissue.
Status
Terminated
Initiated
2013-06-21
Posted
2013-07-20
Terminated
2014-07-15
Reason
The outer box label of the product was labeled with a down revision label which did not include the US labeling requirements, specifically the symbol descriptors and also the "Caution: Federal Law (USA) restricts this device to sale by or on the order of a physician." This product was previously labeled for EU distribution only.
Root cause
Labeling mix-ups
Action
Myoscience representatives visited both affected sites on June 14, 2013 and relabeled the product in person. Neither site had used the product and was unaffected by the relabeling. No formal communication was issued.
Quantity
3 units
Distribution
US distribution including CA and ID.
Lot, serial or model codes
Serial Numbers: 100100, 100101, 100103
510(k) numbers
K123516
PMA numbers
not on file
Product code
GXH
Firm FEI
3009131456
Firm city
Redwood City
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register