Howmedica Osteonics Corp.: Short tibial bearing MK4 - Std, STM, Catalog No. SMMLTB02. Knee prosthesis component.

Recall number
Z-1778-2022
Recalling firm
Howmedica Osteonics Corp.
Product
Short tibial bearing MK4 - Std, STM, Catalog No. SMMLTB02. Knee prosthesis component.
Status
Open, Classified
Initiated
2022-08-24
Posted
2022-09-23
Terminated
not on file
Reason
A MK4 Tibial Bearing was supplied with the intention to mate with the Tibial Metal Casing in-situ. However, the device in-situ was a MK3 Tibial Metal Casing. The supplied MK4 tibial bearing was not exchangeable for the MK3 tibial bearing. The surgeon needed to complete the surgery by re-assembling and re-inserting the extracted MK3 tibial bearing.
Root cause
Under Investigation by firm
Action
On August 24, 2022, the firm sent written notification (Product Field Action) to the single affected customer to document the event. The customer was asked to confirm their receipt of the notice.
Quantity
1 unit
Distribution
Distributed to one customer located in New York.
Lot, serial or model codes
Lot number B24694 (no UDI)
510(k) numbers
["K140898"]
PMA numbers
not on file
Product code
KRQ
Firm FEI
3003070421
Firm city
Mahwah
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register