FDA Medical Device Recalls (openFDA)
GE Healthcare LLC: SIGNA Premier systems
- Recall number
- Z-1767-2026
- Recalling firm
- GE Healthcare LLC
- Product
- SIGNA Premier systems
- Status
- Open, Classified
- Initiated
- 2026-02-27
- Posted
- 2026-04-02
- Terminated
- not on file
- Reason
- Whole-Body MR scanners used to produce images of the inside of the human body to aid diagnosis of disease may have been manufactured/serviced with ferrous (steel) fittings instead of non-ferrous (brass) fittings on the magnet rear and if system is accessed to perform service while the magnet is ramped, the magnetic field could attract the ferrous fittings potentially resulting in personnel injury.
- Root cause
- Nonconforming Material/Component
- Action
- On 2/27/2026, correction notices were mailed to Director of Clinical/Radiology Risk Manager/Hospital Administrators informing them to take the following actions: 1) Ensure all potential servicing personnel in your facility are made aware of this safety notification and the recommended actions. 2) If the device requires service at the rear of the magnet, ensure that all servicing personnel are aware of this potential issue. 3) If servicing personnel have accessed the rear of the magnet with covers removed and identified that ferrous fittings are installed, do NOT attempt to remove or disconnect them. Call your GE HealthCare representative immediately. To help identify the parts: -The incorrect fittings (ferrous) are steel/silver in color -The correct fittings (non-ferrous) are brass/gold in color 4) Complete and return the acknowledgement form electronically via (https://gehealthcare-svc.my.site.com/publicForm/s/?formId=aGjUr000002kiJR) or print, fill out manually, scan and email to Recall.63006@gehealthcare.com If you have any questions or concerns regarding this notification, please contact the firm's HealthCare Service at 1-800-437-1171 or your local Service Representative.
- Quantity
- 41
- Distribution
- Worldwide - US Nationwide distribution in the states of CA, MA, NJ, NY, OH, SD, TX, VA, PR and the countries of Sweden, Switzerland, United Kingdom, Vietnam, China, Norway, Portugal, Spain, France, Germany, Greece, Korea, Republic of
- Lot, serial or model codes
- UDI-DI/GTIN: 00840682135269, 00195278010797. System ID: 415353UMR2, 415723SHMR19, 413253MR3T, 908725MR, 609652MR3T, M48823, 7168453T, 513584MR5, 513636EASTMR, 513636T1MR, 605333MRPRM, 832522CRMR1, 703858PRE2, 804675VA3T, 082427110162, 082427220143, 082427020147, 082427200242, 082427160461, 082427250139, 082427140382, M4216469, 353745MR01, GON5655515, KRMR56474971, KRMR56203791, KRMR55433661, KRMR56344161, KRMR55774051, KRMR55914661, NOMRI26833321, 15974045, GON1882380, ESMRI27056983, ESMRI26960129, A203MR06, 690120MR01, 11524261, VNMRI26792525, VNMRI26592248, VNMRI26809677
- 510(k) numbers
- ["K171128","K183231","K193282"]
- PMA numbers
- not on file
- Product code
- LNH
- Firm FEI
- 3005326970
- Firm city
- Florence
- Firm state
- SC
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28