PerkinElmer Life and Analytical Sciences, Wallac, OY: Specimen Gate Laboratory; Product Number: 5002-0180 Specimen Gate Laboratory is intended for use as a data processing software used in the storage, retrieving, and processing of laboratory data.

Recall number
Z-1761-2017
Recalling firm
PerkinElmer Life and Analytical Sciences, Wallac, OY
Product
Specimen Gate Laboratory; Product Number: 5002-0180 Specimen Gate Laboratory is intended for use as a data processing software used in the storage, retrieving, and processing of laboratory data.
Status
Terminated
Initiated
2017-03-07
Posted
2017-03-13
Terminated
2021-05-10
Reason
Potential for errors in patient results generated by the Specimen Gate Laboratory Software.
Root cause
Software design
Action
PerkinElmer Life and Analytical Sciences sent an Urgent Medical Device Correction letter dated March 7, 2017, to all affected consignees. The letter contains the following instructions: Until PerkinElmer's corrective measures are completed, we request our customers to close the Specimen Gate Laboratory Result Viewer between viewing and especially between manually calculating assay results. Closing and reopening the Result Viewer between processing the results from different assays effectively prevents the described issues from occurring. We also request our customers to follow unreported specimens with available tools to avoid any delay in reporting screening results. The available tools will vary depending on their process or the LIMS system that is used for activities associated with management of patient demographics, patient reports and the overall management of data generated from newborn screening testing. The final corrective action will involve an installation update to the software. PerkinElmer will deploy an installation update that will prevent the described issue when assays are calculated during the same Result Viewer session without closing it between the processing of different assays. Once the installation update is available, Software Services will contact you and arrange for the deployment. The installation update will be provided to you free of charge. Customers were instructed to complete the enclosed response form and return it by fax to 1-330-825-8520 (US) =358 2 2678 357 (OUS). For questions regarding this recall call 317-418-1735.
Quantity
29
Distribution
Worldwide Distribution - US including AL, AZ, AR, CA, CO, CT, FL, GA, IL, IN, KS, KY, LA, MD, MI, MO, NV, OH, OK, OR, TN, TX, VA; and Internationally to Canada and Brazil
Lot, serial or model codes
Software Version: 1.3 Rev2, 1.7, 1.8, 1.9, 1.9 Rev2
510(k) numbers
not on file
PMA numbers
not on file
Product code
JQP
Firm FEI
3002808498
Firm city
Turku
Firm state
not on file
Firm country
Finland

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register