Ge Healthcare, Llc: GE Healthcare, Seno Advantage 2.2 workstations with software version 22_01 and 22_02 Seno Advantage is a medical image review station that allows easy selection, processing, filming and media interchange of multi-modality images from a variety of diagnosis imaging systems. When interpreted by a trained physician, mammographic images displayed on the high-resolution monitors may be used as an element for diagnosis. Furthermore filmed images from all modalities may also be used as an element for diagnosis.

Recall number
Z-1747-2010
Recalling firm
Ge Healthcare, Llc
Product
GE Healthcare, Seno Advantage 2.2 workstations with software version 22_01 and 22_02 Seno Advantage is a medical image review station that allows easy selection, processing, filming and media interchange of multi-modality images from a variety of diagnosis imaging systems. When interpreted by a trained physician, mammographic images displayed on the high-resolution monitors may be used as an element for diagnosis. Furthermore filmed images from all modalities may also be used as an element for diagnosis.
Status
Terminated
Initiated
2010-02-09
Posted
2010-06-02
Terminated
2011-12-18
Reason
GE Healthcare has recently become aware of inaccurate Scale factor annotation provided by SenoAdvantage 2.2 workstations that may impact patient safety. When using the Print Image or Print Screen feature on SenoAdvantage 2.2, the Scale factor annotation printed on hard copy images is inaccurate. Secondary Capture (SCPT) images displayed on the screen and Secondary Capture hard copy images may a
Root cause
Software design
Action
The firm, GE Healthcare, sent an "Urgent Medical Device Correction" letter dated February 8, 2010 to customers. The letter described the Safety Issue, Affected product details, Safety Instructions to prevent further inaccurate scale factor annotation, Product Correction and Contact Information. GE Healthcare will correct all affected workstations by providing a software upgrade and an updated Operator Manual. The GE service representative will contact customer to arrange for this correction. If you have any questions or concerns regarding this notification, please call one of the following phone numbers: United States: 800-437-1171, Japan: -0120-055-919. For other countries, please contact your local GE Healthcare Service Representative.
Quantity
73 devices
Distribution
Worldwide distribution: USA including states of CA, FL, GA, MD, NJ, NM, NY, NC, OK, TX and countries of THAILAND, TAIWAN, SWEDEN SPAIN, SAUDI ARABIA, RUSSIAN FEDERATION, PORTUGAL, PAKISTAN, MEXICO, JAPAN, ITALY,GREECE, FRANCE, COLOMBIA, CHILE, CANADA, BELGIUM, and BAHRAIN.
Lot, serial or model codes
SERIAL NUMBER 000000A9316002 000000A9316008 000000F9314002 000000F9314013 000000Q9309007 000000Q9309012 000000Q9309013 000000Q9309015 000000W9321002 000000W9321004 000000A9316004 000000A9316014 000000A9316015 000000A9345002 000000A9345003 000000A9345007 000000A9345009 000000A9345019 000000A9345020 000000F9314005 000000F9314010 000000F9314014 000000Q9309002 000000Q9309003 000000Q9309006 000000W9321003 000000P9300003 000000A9316012 000000P9300002 000000F9314011 000000W9321001 000000W9321007 000000W9321008 000000W9321010 000000W9321006 000000A9316006 000000A9316003 000000A9316005 000000P9300001 000000Q9309010 000000Q9309014 000000A9345015 000000Q9309005 000000A9316011 000000A9345012 000000A9345013 000000A9345014 000000A9345016 000000A9345018 000000F9314004 000000Q9309001 00000D10011003 00000D10011006 000000A9316001 000000F9314008 000000F9314012 000000P9300004 000000W9321009 000000A9316010 000000F9314007 000000W9321005 000000A9316007 000000A9316009 000000F9314015 000000F9314006 000000Q9356005 00000ZA9260001 000000Q9309008 000000A9345001 000000A9345017 000000F9314001 000000Q9309004 000000Q9356001
510(k) numbers
["K033400"]
PMA numbers
not on file
Product code
LLZ
Firm FEI
2126677
Firm city
Waukesha
Firm state
WI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register