Skip to the content

FDA Medical Device Recalls (openFDA)

Inari Medical - Oak Canyon: Artix MT Thrombectomy Device, REF: 32-102

Recall number
Z-1728-2025
Recalling firm
Inari Medical - Oak Canyon
Product
Artix MT Thrombectomy Device, REF: 32-102
Status
Open, Classified
Initiated
2025-04-08
Posted
2025-05-06
Terminated
not on file
Reason
Thrombectomy device specifies vessel range of 2.5-6 mm on the pouch, carton, and Instructions for Use rather than the correct vessel range of 3-6 mm, which could incrementally increase the radial force on the artery, which has the potential of vascular complications.
Root cause
Labeling Change Control
Action
On 4/8/2025, correction and removal notices were distributed to customers who were asked to do the following: 1) Review notice and the accompanying copy of the device IFU (IU-01087 Rev. C) and disseminate it to any applicable personnel and device users at your facility and network to ensure they are aware of the unused device replacement. 2) Notice should be shared with any organization where affected devices may have been transferred. 3) Complete and return the acknowledgement reply form via email to If you have any questions, please contact your local sales representative, or the firm's Customer Service at 877-923-4747, or Quality department at QA@inarimedical.com.
Quantity
45
Distribution
US Nationwide distribution in the states of FL, IL, WI, KY, TX, CA, OH, GA, AK, MI, PA, WA, CO, OH.
Lot, serial or model codes
UDI-DI: 00810123710584, Lot: 24100082, Expiration: 14 April 2025. IFU: IU-01087 Rev. B
510(k) numbers
K241894
PMA numbers
not on file
Product code
QEW
Firm FEI
3020347218
Firm city
Irvine
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register