Skip to the content

FDA Medical Device Recalls (openFDA)

Integra LifeSciences Corp. (NeuroSciences): CODMAN Craniotomy Kit, Disposable Perforator Cranio-blade Wire Pass Drill, 14mm. Cranial perforator kit.

Recall number
Z-1717-2025
Recalling firm
Integra LifeSciences Corp. (NeuroSciences)
Product
CODMAN Craniotomy Kit, Disposable Perforator Cranio-blade Wire Pass Drill, 14mm. Cranial perforator kit.
Status
Open, Classified
Initiated
2025-04-11
Posted
2025-05-06
Terminated
not on file
Reason
Inadequate weld that can potentially cause the product to disassemble.
Root cause
Process control
Action
An URGENT: VOLUNTARY MEDICAL DEVICE RECALL notification, dated 4/11/25, was distributed to consignees via mail and email. Consignees are being asked to check their inventory for affected devices and immediately quarantine any product identified. Consignees are asked to complete the provided Acknowledgement Form and return it via email to FCA3@integralife.com or by fax to 1-609-750-4220; all consignees are to complete and return the form. Integra asks that the recall notification be forwarded to all users of the devices and to customers who received product that was further distributed. Distributors are to collect completed Acknowledgement Forms from their customers and provide them to the firm. Consignees with any questions can contact Customer Service at 1-800-654-2873 or by email at custsvcnj@integralife.com.
Quantity
1,840 units
Distribution
Worldwide distribution - US Nationwide including Puerto Rico and the countries of Argentina, Australia, Austria, Bahamas, Barbados, Belgium, Brazil, Canada, Chile, Colombia, Cyprus, Czechia, Dominican Republic, Ecuador, El Salvador, France, French Guiana, Germany, Ghana, Greece, Hong Kong, India, Ireland, Israel, Italy, Japan, Jordan, Kenya, Korea (the Republic of), Lebanon, Lithuania, Malaysia, Morocco, Nepal, Netherlands, New Zealand, Pakistan, Poland, Portugal, Qatar, Saudi Arabia, Singapore, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Thailand, Trinidad and Tobago, Tunisia, Turkey, United Arab Emirates, United Kingdom.
Lot, serial or model codes
Model No. 261230; UDI: 10381780513629; Lot No. Lot Number, 5984530, 7062002, 7062006, 7062013, 7062014, 7255930, 7255931, 7255932, 7255933, 7255934, 7255935, 7255936, 7255937, 7269522, 7269523, 7269524, 7269525, 7282944, 7282945, 7282946, 7282948, 7282949, 7282950, 7282951, 7282952, 7282953, 7293408, 7293409, 7293410, 7293411, 7293412, 7293413, 7293414, 7322892, 7322893, 7322894, 7322895, 7322896, 7322897, 7322898, 7332664, 7332665, 7332666, 7332667, 7332668, 7348744, 7348745, 7348746, 7348747, 7348748, 7371080, 7371081, 7379573, 7379576, 7379577, 7379578.
510(k) numbers
["K073336"]
PMA numbers
not on file
Product code
HBF
Firm FEI
3003418325
Firm city
Princeton
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register