Skip to the content

FDA Medical Device Recalls (openFDA)

Sysmex America, Inc.: The CF-70 instrument (product code: LXG; Regulation Number: 862.2050) receives magazines containing stained blood smears from SP-50, selects only smears that require analysis, and transports them to the DI-60 for further analysis. After completing the analysis on DI-60 the smears are placed back in magazines.

Recall number
Z-1703-2020
Recalling firm
Sysmex America, Inc.
Product
The CF-70 instrument (product code: LXG; Regulation Number: 862.2050) receives magazines containing stained blood smears from SP-50, selects only smears that require analysis, and transports them to the DI-60 for further analysis. After completing the analysis on DI-60 the smears are placed back in magazines.
Status
Terminated
Initiated
2020-03-09
Posted
not on file
Terminated
2021-03-24
Reason
Software mismatch-When the software versions between the SP-50 and CF-70 are not matched, and an error condition occurs requiring the operator to open the cover door of the CF-70 to remove slides or slide magazines from the CF-70, the operation of the CF-70 does not halt.
Root cause
Software design
Action
Sysmex America, Inc. notified customers on 03/09/2020 via "Product Notification SP-50 and CF-70 Software Mismatch Field Correction" letter. The recall letter identified the affected product and model numbers. The customers were asked to verify the software version. If the versions did not match, a service visit is required for the correction of the software mismatch.
Quantity
93 units.
Distribution
USA nationwide distribution in the states of AK, AR, AZ, CA, CO, CT, DE, FL, IL, KY, LA, MA, MD, MI, MO, MS, MT, NC, NJ, NM, NY, OH, OK, OR, PA, TN, TZ, UT and WA..
Lot, serial or model codes
Catalog No. AH699752.
510(k) numbers
not on file
PMA numbers
not on file
Product code
LXG
Firm FEI
3009711478
Firm city
Lincolnshire
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
3e6df46828c1
Size
288213574 bytes
Published
2026-10-06
Read
2026-10-07T03:20:25.099Z
Source file
Source file retained · verify
Refresh due
2026-10-14

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.